4.1.3. Setting the Display Freeze Function
1. Press
MENU
(switch to the setting mode).
2. Select
Timer, Event action
> Event action.
3. Set
the
Logic box number
, Event (set to Remote), and Remote number.
4. In
the
Action
box, press the Freeze soft key.
Setup Screen Example for the Display Freeze Function (DX1000(N) Example)
4.1.4. Operation While the Display is frozen
If the DX is in a state in which it can switch to the basic setting mode, the DX can
switch to the mode even while the display is frozen.
¾
The display freeze function is released when switching back to the operation
screen from the basic setting mode.
If the power turns OFF and then back ON such as due to a power failure while the
display is frozen (closed state) through remote control, the display freeze function is
released even if there is no change in the remote state (remains at the closed
state).
The automatic display revert function does not operate.
¾
The automatic revert timer is restarted when the display freeze function is
released.