3. Accuracy and Display Resolution

3.1. Measurement and Display Accuracy (Digital Display)

Computing method of the display value accuracy

¾

Input

voltage

X

¾

Lower limit of display span VL

Upper limit of display span VU

¾

Scale lower limit

SL

Scale upper limit

SU

¾

Digital display value

Y

1) Converting equation for the input voltage (setting)

• Y = 10

(SU-SL)*(X-VL)/(VU-VL)+SL

2) Measurement accuracy of input voltage X (uses the negative side due to the

LOG characteristic)
• X

err

=X – (measurement accuracy of the voltage range)

3) Hardware error of the display value

• Y’= Y – Y

err

• The difference from the true value Y is hardware error Y’ when display

value Y

err

corresponds to input voltage X

err

.

4) Display accuracy of the mantissa

• Display accuracy = Y’* 1.1 + 1 digit