Accuracy and display resolution – Yokogawa Button Operated DX2000 User Manual
Page 11
10
3. Accuracy and Display Resolution
3.1. Measurement and Display Accuracy (Digital Display)
Computing method of the display value accuracy
¾
Input
voltage
X
¾
Lower limit of display span VL
Upper limit of display span VU
¾
Scale lower limit
SL
Scale upper limit
SU
¾
Digital display value
Y
1) Converting equation for the input voltage (setting)
• Y = 10
(SU-SL)*(X-VL)/(VU-VL)+SL
2) Measurement accuracy of input voltage X (uses the negative side due to the
LOG characteristic)
• X
err
=X – (measurement accuracy of the voltage range)
3) Hardware error of the display value
• Y’= Y – Y
err
• The difference from the true value Y is hardware error Y’ when display
value Y
err
corresponds to input voltage X
err
.
4) Display accuracy of the mantissa
• Display accuracy = Y’* 1.1 + 1 digit