Event action functions that do not operate while the display is frozen
¾
See
Table
4-1.
¾
If a level action such as Memory start/stop is executed due to a level event
such as remote while the display is frozen, the DX operation may become
mismatched with the event state.
• Example) If Memory start/stop using remote is specified and the remote
control input is switched from closed to open while the display is frozen, the
memory stop operation is not executed. If display freeze is released in this
condition, the remote control input terminal will be open even though the
memory is in a start condition.
Table 4-1 Event Action Functions That Do Not Operate While the Display Is Frozen
Action
Note
Memory start/stop
Does not operate while the display is frozen. Operates after the
display freeze is released.
Memory start
Does not operate while the display is frozen. Operates after the
display freeze is released.
Memory stop
Does not operate while the display is frozen. Operates after the
display freeze is released.
Event trigger
Does not operate while the display is frozen. Operates after the
display freeze is released.
However, the action does not operate if the event is set to timer or
match timer.
Save display data
Does not operate while the display is frozen. Does not operate even
after the display freeze is released.
Save event data
Does not operate while the display is frozen. Does not operate even
after the display freeze is released.
Manual sample
Does not operate while the display is frozen. Does not operate even
after the display freeze is released.