2.10. Communication Commands
2.10.1. Setting the LOG Range
Setting mode setting
SR p1,p2,p3,p4,p5,p6,p7,p8,p9
¾
p1: Channel number (001 to 048)
¾
p2: Setting
type
(Log)
¾
p3: Measurement range (20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V, or 50 V)
¾
p4: Span
lower
limit
¾
p5: Span upper limit p4 < p5
¾
p6: Lower limit of exponent (-15 to 15)
¾
p7: Upper limit of exponent (-15 to 15) p6 < p7 and (p7-p6) ≤ 15
¾
p8: Decimal place (fixed to 0) * Decimal place of the exponent
¾
p9: Unit (up to 6 characters)
2.10.2. Specifying the Number of Displayed Mantissa Digits
Basic setting mode setting
QA p1
¾
p1: Number of displayed digits (2: 2 digits or 3: 3 digits)
2.11. Manual Sample Function
Stored data when a channel set to LOG is assigned
¾
The data is stored in the same format as with the digital value display
(mantissa + exponent).
Events that cause the manual sampled data to be divided
¾
If the number of displayed mantissa digits of the LOG input range is changed,
the data file is divided the next time manual sample is executed.