If it is more convenient, the user may make solutions from Sodium Chloride (NaCl or common

table salt) with the help of the following relationship table. This table is derived from the IEC norm

60746-3.

Table 11-3. NaCl values at 25 °C

Weight %

mg/kg

Conductivity

0.001

10

21.4 µS/cm

0.003

30

64.0 µS/cm

0.005

50

106 µS/cm

0.01

100

210 µS/cm

0.03

300

617 µS/cm

0.05

500

1.03 mS/cm

0.1

1000

1.99 mS/cm

0.3

3000

5.69 mS/cm

0.5

5000

9.48 mS/cm

1

10000

17.6 mS/cm

3

30000

48.6 mS/cm

5

50000

81.0 mS/cm

10

100000

140 mS/cm

Note!

For resistivity measurement the standard resistivity units of the calibration solution can be calcu-

lated as follows:

R = 1000/G (kΩ•cm if G = µS/cm)

Example: 0.001% weight

R = 1000/21.4 = 46.7 kΩ•cm