Yokogawa EXAxt SC450 4-Wire Analyzer for Conductivity/Resistivity User Manual
Page 22

14
IM 12D08N05-01E
Extension cable may be purchased in bulk
quantities or in pre-finished lengths. In the case
of bulk quantities cut to length, then it is neces-
sary to terminate the cable as shown below.
Termination procedure for WF10 cable.
1. Slide 3 cm of heat shrink tube (9 x 1.5)
over the cable end to be terminated.
2. Strip 9 cm of the outer (black) insulating
material, taking care not to cut or damage
internal cores.
3. Remove loose copper screening, and cut
off the cotton packing threads as short as
possible.
4. Strip insulation from the last 3 cm of the
brown, and the white coaxial cores.
5. Extract the coaxial cores from the
braid, and trim off the black (low-noise)
screening material as short as possible.
6. Insulate the overall screen and the 2
coaxial screens with suitable plastic
tubing.
7. Strip and terminate all ends with suitable
(crimp) terminals and identify with
numbers as shown.
8. Finally shrink the overall heat shrink tube
into position.
Figure 3-11.a.
3 cm
heat shrink
9 cm
remove insulation
Figure 3-11.b.
3 cm
cotton threads
Figure 3-11.c.
11
12
17
15
16
13
14