<2. Notes on Handling>
2-9
IM 01C50E01-01EN
WARNING
• Potential electrostatic charging hazard -
secure distance of 100mm from antenna.
• Take care not to generate mechanical
sparking when access to the instrument and
peripheral devices in a hazardous location.
2.8 EMC Conformity Standards
EN61326-1 Class A, Table 2 (For use in
industrial locations), EN61326-2-3
CAUTION
This instrument is a Class A product, and it is
designed for use in the industrial environment.
Please use this instrument in the industrial
environment only.
WARNING
The potential for discharging static electricity
should be minimized in mounting this transmitter.
For example, do not generate static electricity
as the surface of a battery pack or housing is
rubbed with cloth.
2.9 Low Voltage Directive
Applicable standard:
EN61010-1, EN61010-2-030
(1) Pollution Degree 2
"Pollution degree" describes the degree to
which a solid, liquid, or gas which deteriorates
dielectric strength or surface resistivity is
adhering. " 2 " applies to normal indoor
atmosphere. Normally, only non-conductive
pollution occurs. Occasionally, however,
temporary conductivity caused by condensation
must be expected.
(2) Installation Category I
(Anticipated transient overvoltage 330 V)
"Overvoltage category (Installation category)"
describes a number which defines a transient
overvoltage condition. It implies the regulation
for impulse withstand voltage. " I " applies to
electrical equipment which is supplied from the
circuit when appropriate transient overvoltage
control means (interfaces) are provided.
2.10 Regulatory Compliance
for Radio and
Telecommunication
Please confirm that a installation region fulfills
a standards, require additional regulatory
information and approvals, contact to
Yokogawa Electric Corporation.
2.11 Radio and
Telecommunications
Terminal Equipment
Directive (R&TTE)
We, Yokogawa Electric Corporation hereby declare
that this equipment,model YTA510 series is in
compliance with the essential requirements and
other relevant provisions of Directive 1999/5/EC.
The CE declaration of conformity for R&TTE for this
product can be found at
http://www.yokogawa.com/fld/
2.12 FCC compliance
This equipment contains transmitter module FCC
ID: SGJ-WFC001.
This device complies with Part 15 of FCCRules.
Operation is subject to the following two conditions:
(1) this device may not cause interference, and (2)
this devicemust accept any interference, including
interference that may cause undesired operation of
this device.