<11. General Specifications>
11-2
IM 01C27F01-01EN
Working Pressure Limits (Silicone oil):
Maximum Pressure Limits
Pressure
Capsule
EJX510B
EJX530B
A
200 kPa abs (29 psia)
200 kPa (29 psig)
B
2 MPa abs (290 psia)
2 MPa (290 psig)
C
10 MPa abs (1450 psia) 10 MPa (1450 psig)
D
50 MPa abs (7200 psia) 50 MPa (7200 psig)
Minimum Pressure Limit
See graph below
-40
(-40)
0
(32)
40
(104)
80
(176)
120 (248)
0.1(0.75)
0.13(1)
1(7.5)
100(750)
10(75)
2.7(20)
0.01(0.075)
0.013(0.1)
85 (185)
(mmHg abs)
Working
pressure
kPa abs
Applicable range
Process temperature °C ( °F)
F01E.ai
Atmospheric
pressure
-40
(-40)
0
(32)
40
(104)
80
(176)
120
(248)
1(0.14)
2.7(0.38)
10(1.4)
(psia)
100(14.5)
Process temperature °C ( °F)
Working
pressure
kPa abs
Applicable range
F02E.ai
Maximum Over Pressure
Pressure
Capsule
EJX510B
EJX530B
A and B
4 MPa abs (580 psia)
4 MPa (580 psig)
C
20 MPa abs (2900 psia) 20 MPa (2900 psig)
D
75 MPa abs
(10800 psia)
75 MPa
(10800 psig)
EMC Conformity Standards
EN61326-1 Class A, Table 2 (For use in
industrial locations), EN61326-2-3
R&TTE Conformity Standards
ETSI EN 300 328, ETSI EN 301 489-1,
ETSI EN 301 489-17, EN61010-1,
EN61010-2-030, EN62311
• Indoor/Outdoor use
Regulation Conformity of the Wireless Module
• FCC Approval
• IC Approval
Safety Requirement Standards
EN61010-1, EN61010-2-030
• Altitude of installation site:
Max. 2,000 m above sea level
• Installation category: I
(Anticipated transient overvoltage 330 V)
• Pollution degree: 2
• Indoor/Outdoor use
Connections:
Refer to “Model And Suffix Codes.”
Wetted Parts Materials:
Diaphragm, Process Connector
Refer to “Model And Suffix Codes.”
Non-wetted Parts Materials:
Housing
Low copper cast aluminum alloy with
polyurethane, mint-green paint (Munsell
5.6BG 3.3/2.9 or its equivalent)
Degrees of Protection
IP66/IP67, NEMA4X
Pipe
Polypropylene
Cover O-rings
Buna-N
Name plate and tag
316 SST tag plate wired onto transmitter
Fill Fluid
Silicone, fluorinated oil (optional)
Weight:
Capsule A, B and C: 3.2kg (7.1 lb) *
Capsule D: 3.4kg (7.4 lb) *
*: Without battery pack and mounting bracket.