Yokogawa EJX530B User Manual
Page 32

<7. Operation>
7-4
IM 01C27F01-01EN
■ Connecting to a field wireless network
The action after installing the battery pack varies
depending on the silence setting.
Mounting the battery pack automatically starts a
search for the field wireless network. When the
Field Wireless Integrated Gateway is found, the
instrument enters the join status.
When the Field Wireless Integrated Gateway is not
found and a specified time has elapsed, a cycle of a
one-hour pause and six-minutes search is repeated
until the instrument can join the field wireless
network.
For details on the procedure to switch to silence
mode, refer to subsection 8.3.13 “Switching to
Silence Mode.”
F0704.ai
Mounting battery pack
Boot
Infrared communication
Pause (b)
Search failure for
the specified time
Infrared
communication
1 hour
passed
Search failure
for 6 minutes
Ready 2:
Searching (b)
(Connect)
(Connect)
Confirm connecting
status: Join (c)
(Publish)
Operation (d)
Deep sleep
setting
(Disconnect)
(Disconnect)
Ready 1:
Searching (b)
Deep sleep (a)
Infrared
communication
Figure 7.4
Wireless Connection Process
(a) Deep sleep
F0705.ai
(b) Ready and pause
F0706.ai
(c) Confirm connecting status
F0707.ai
(d) Join
F0708.ai