General specifications, 1 standard specifications, General specifications -1 – Yokogawa EJX430B User Manual
Page 92: Standard specifications -1, Ejx310b, Ejx430b

<11. General Specifications>
11-1
IM 01C27B01-01EN
11. General Specifications
11.1 Standard Specifications
Communication protocol:
ISA100.11a protocol
Data rate:
250 kbps
Frequency:
2400 - 2483.5 MHz license free ISM band
Radio security:
AES 128 bit codified
RF Transmitter power:
Max. 11.6 dBm (fixed)
Antenna:
+2 dBi Omni directional monopole type
Span and range limits:
EJX110B
Measurement
Span/Range
kPa
inH
2
O(/D1)
mbar(/D3) mmH
2
O(/D4)
F
Span
0.1 to 5
0.4 to 20
1 to 50
10 to 500
Range
-5 to 5
-20 to 20
-50 to 50
-500 to 500
L
Span
0.1 to 10
0.4 to 40
1 to 100
10 to 1000
Range
-10 to 10
-40 to 40
-100 to 100 -1000 to 1000
M
Span
0.5 to 100
2 to 400
5 to 1000
50 to 10000
Range -100 to 100 -400 to 400 -1000 to 1000
-10000 to
10000
H
Span
2.5 to 500
10 to 2000
25 to 5000
0.025 to 5
kgf/cm
2
Range -500 to 500 -2000 to 2000 -5000 to 5000
-5 to 5
kgf/cm
2
V
Span
0.07 to 14
MPa
10 to 2000 psi 0.7 to 140 bar
0.7 to 140
kgf/cm
2
Range
-0.5 to 14
MPa
-71 to 2000
psi
-5 to 140 bar
-5 to 140
kgf/cm
2
EJX310B
Measurement
Span/Range
kPa abs
psi abs(/D1)
mbar abs
(/D3)
mmHg abs
(/D4)
L
Span
0.5 to 10
0.15 to 2.95
inHg
5 to 100
3.8 to 75
Range
0 to 10
0 to 2.95 inHg
0 to 100
0 to 75
M
Span
1.3 to 130
0.39 to 38
inHg
13 to 1300
9.8 to 970
Range
0 to 130
0 to 38 inHg
0 to 1300
0 to 970
A
Span
0.0175 to
3.5 MPa
2.5 to 500
0.175 to 35
bar
0.175 to 35
kgf/cm
2
Range
0 to 3.5
MPa
0 to 500
0 to 35 bar
0 to 35
kgf/cm
2
B
Span
0.08 to 16
MPa
12 to 2300 0.8 to 160 bar
0.8 to 160
kgf/cm
2
Range 0 to 16 MPa
0 to 2300
0 to 160 bar
0 to 160
kgf/cm
2
EJX430B
Measurement
Span/Range
MPa
psi(/D1)
bar(/D3)
kgf/cm
2
(/D4)
H
Span
2.5 to 500
kPa
10 to 2000
inH
2
O
0.025 to 5
0.025 to 5
Range
-100 to 500
kPa
-400 to 2000
inH
2
O
-1 to 5
-1 to 5
A
Span
0.0175 to
3.5
2.5 to 500
0.175 to 35
0.175 to 35
Range -0.1 to 3.5 -14.5 to 500
-1 to 35
-1 to 35
B
Span
0.08 to 16
12 to 2300
0.8 to 160
0.8 to 160
Range
-0.1 to 16 -14.5 to 2300
-1 to 160
-1 to 160
Performance specifications:
Refer to GS01C27B01-01EN.
Update Time
Measurement
mode
Differential
pressure
Pressure
Continuous
100 ms
100 ms
Intermittent
0.5 to 3600 s
selectable
0.5 to 3600 s
selectable
For amplifier housing code 7:
The transmitter shifts to the countinuous mode
when the update time is set to 1 second.
Minimum update time is 1 second.
For amplifier housing code 8 and 9:
The transmitter shifts to the countinuous mode
when the update time is set to 0.5 second.
Power Supply Specifications
Battery:
Use the dedicated battery pack.
Rated voltage: 7.2 V
Rated capacity: 19 Ah
Output:
ISA100.11a protocol
Output mode (EJX110B):
linear or square root
Ambient Temperature Limits:
-40 to 85°C (-40 to 185°F)
-30 to 80°C (-22 to 176°F) LCD visible range
Process Temperature Limits:
-40 to 120°C (-40 to 248°F)
– Except EJX310B L capsule
-40 to 100°C (-40 to 212°F)
– EJX310B L capsule
Ambient Humidity Limits:
0 to 100% RH