beautypg.com

General specifications, 1 standard specifications, General specifications -1 – Yokogawa EJA530A User Manual

Page 57: 1 standard specifications -1

background image

IM 01C21F01-01E

10-1

10. GENERAL SPECIFICATIONS

10. GENERAL SPECIFICATIONS

10.1

Standard Specifications

Refer to IM 01C22T02-01E for F

OUNDATION

Fieldbus communication type and IM 01C22T03-
00E for PROFIBUS PA communication type
marked with “

᭛”.

Performance Specifications

See General Specifications sheet, GS 01C21F01-E.

Functional Specifications

Span & Range Limits

EJA510A and EJA530A:

A

B

Measurement

Span and

Range

Span

Range

Span

Range

psi (/D1)

1.45 to 29

0 to 29

14.5 to 290

0 to 290

bar (/D3)

0.1 to 2

kgf/cm

2

(/D4)

0.1 to 2

0 to 2

1 to 20

0 to 20

0 to 2

1 to 20

0 to 20

MPa

10 to 200

kPa

0 to 200

kPa

0.1 to 2

0 to 2

72.5 to 1450

0.5 to 10

0 to 10

C

Span

Range

0 to 1450

5 to 100

0 to 100

5 to 100

0 to 100

720 to 7200

5 to 50

0 to 50

D

Span

Range

0 to 7200

50 to 500

0 to 500

50 to 500

0 to 500

T1001.EPS

Values in absolute for EJA510A.

Zero Adjustment Limits:

Zero can be fully elevated or suppressed, within the
Lower and Upper Range Limits of the capsule.

External Zero Adjustment “

”:

External zero is continuously adjustable with 0.01%
incremental resolution of span. Span may be
adjusted locally using the digital indicator with range
switch.

Output “

”:

Two wire 4 to 20 mA DC output with digital commu-
nications, linear or square root programmable.
BRAIN or HART FSK protocol are superimposed on
the 4 to 20 mA signal.

Failure Alarm:

Output status at CPU failure and hardware error;

Up-scale: 110%, 21.6 mA DC or more(standard)
Down-scale: -5%, 3.2 mA DC or less

-2.5%, 3.6 mA DC or less(Optional

code /F1)

Note: Applicable for Output signal code D and E

Damping Time Constant (1st order):

The sum of the amplifier and capsule damping
time constant must be used for the overall time
constant. Amp damping time constant is adjustable
from 0.2 to 64 seconds.

Time Constant (approx. sec)

0.2

A, B, C, and D

T1002.EPS

Capsule (Silicone Oil)

Ambient Temperature Limits:

* Safety approval codes may affect limits.

–40 to 85°C (–40 to 185°F),
–30 to 80°C (–22 to 176°F) with LCD Display

Process Temperature Limits:

* Safety approval codes may affect limits.

–40 to 120°C (–40 to 248°F)

Maximum Overpressure:

A

4 MPa abs{580 psia}

4 MPa {580 psig}

B

4 MPa abs{580 psia}

4 MPa {580 psig}

C

20 MPa abs{2900 psia}

20 MPa {2900 psig}

D

60 MPa abs{8500 psia}

60 MPa {8500 psig}

EJA530A

EJA510A

T1003.EPS

Capsule

Capsule

Working Pressure Limits (Silicone Oil)

Maximum Pressure Limit:

A

200 kPa abs{29 psia}

200 kPa {29 psig}

B

2 MPa abs{290 psia}

2 MPa {290 psig}

C

10 MPa abs{1450 psia} 10 MPa {1450 psig}

D

50 MPa abs{7200 psia} 50 MPa {7200 psig}

EJA530A

EJA510A

T1004.EPS

Capsule

Capsule

Minimum Pressure Limit:

EJA510A: 0.013 kPa abs
EJA530A: Lower limit of measurement range

Installation

Supply & Load Requirements “

”:

* Safety approvals can affect electrical requirements.

See Section 6.6, ‘Power Supply Voltage and Load
Resistance.’

Supply Voltage “

”:

10.5 to 42 V DC for general use and flameproof type
10.5 to 32 V DC for lightning protector (Optional code /A)
10.5 to 30 V DC for intrinsically safe, Type n,
nonincendive, or non-sparking type
Minimum voltage limited at 16.4 V DC for digital
communications, BRAIN and HART