General specifications, 1 standard specifications, General specifications -1 – Yokogawa EJA530A User Manual
Page 57: 1 standard specifications -1
IM 01C21F01-01E
10-1
10. GENERAL SPECIFICATIONS
10. GENERAL SPECIFICATIONS
10.1
Standard Specifications
Refer to IM 01C22T02-01E for F
OUNDATION
Fieldbus communication type and IM 01C22T03-
00E for PROFIBUS PA communication type
marked with “
᭛”.
᭹ Performance Specifications
See General Specifications sheet, GS 01C21F01-E.
᭹ Functional Specifications
Span & Range Limits
EJA510A and EJA530A:
A
B
Measurement
Span and
Range
Span
Range
Span
Range
psi (/D1)
1.45 to 29
0 to 29
14.5 to 290
0 to 290
bar (/D3)
0.1 to 2
kgf/cm
2
(/D4)
0.1 to 2
0 to 2
1 to 20
0 to 20
0 to 2
1 to 20
0 to 20
MPa
10 to 200
kPa
0 to 200
kPa
0.1 to 2
0 to 2
72.5 to 1450
0.5 to 10
0 to 10
C
Span
Range
0 to 1450
5 to 100
0 to 100
5 to 100
0 to 100
720 to 7200
5 to 50
0 to 50
D
Span
Range
0 to 7200
50 to 500
0 to 500
50 to 500
0 to 500
T1001.EPS
Values in absolute for EJA510A.
Zero Adjustment Limits:
Zero can be fully elevated or suppressed, within the
Lower and Upper Range Limits of the capsule.
External Zero Adjustment “
᭛”:
External zero is continuously adjustable with 0.01%
incremental resolution of span. Span may be
adjusted locally using the digital indicator with range
switch.
Output “
᭛”:
Two wire 4 to 20 mA DC output with digital commu-
nications, linear or square root programmable.
BRAIN or HART FSK protocol are superimposed on
the 4 to 20 mA signal.
Failure Alarm:
Output status at CPU failure and hardware error;
Up-scale: 110%, 21.6 mA DC or more(standard)
Down-scale: -5%, 3.2 mA DC or less
-2.5%, 3.6 mA DC or less(Optional
code /F1)
Note: Applicable for Output signal code D and E
Damping Time Constant (1st order):
The sum of the amplifier and capsule damping
time constant must be used for the overall time
constant. Amp damping time constant is adjustable
from 0.2 to 64 seconds.
Time Constant (approx. sec)
0.2
A, B, C, and D
T1002.EPS
Capsule (Silicone Oil)
Ambient Temperature Limits:
* Safety approval codes may affect limits.
–40 to 85°C (–40 to 185°F),
–30 to 80°C (–22 to 176°F) with LCD Display
Process Temperature Limits:
* Safety approval codes may affect limits.
–40 to 120°C (–40 to 248°F)
Maximum Overpressure:
A
4 MPa abs{580 psia}
4 MPa {580 psig}
B
4 MPa abs{580 psia}
4 MPa {580 psig}
C
20 MPa abs{2900 psia}
20 MPa {2900 psig}
D
60 MPa abs{8500 psia}
60 MPa {8500 psig}
EJA530A
EJA510A
T1003.EPS
Capsule
Capsule
Working Pressure Limits (Silicone Oil)
Maximum Pressure Limit:
A
200 kPa abs{29 psia}
200 kPa {29 psig}
B
2 MPa abs{290 psia}
2 MPa {290 psig}
C
10 MPa abs{1450 psia} 10 MPa {1450 psig}
D
50 MPa abs{7200 psia} 50 MPa {7200 psig}
EJA530A
EJA510A
T1004.EPS
Capsule
Capsule
Minimum Pressure Limit:
EJA510A: 0.013 kPa abs
EJA530A: Lower limit of measurement range
᭹ Installation
Supply & Load Requirements “
᭛”:
* Safety approvals can affect electrical requirements.
See Section 6.6, ‘Power Supply Voltage and Load
Resistance.’
Supply Voltage “
᭛”:
10.5 to 42 V DC for general use and flameproof type
10.5 to 32 V DC for lightning protector (Optional code /A)
10.5 to 30 V DC for intrinsically safe, Type n,
nonincendive, or non-sparking type
Minimum voltage limited at 16.4 V DC for digital
communications, BRAIN and HART