7) unit setup for displayed temperature, 7) unit setup for displayed temperature -11 – Yokogawa EJA530A User Manual
Page 42
IM 01C21F01-01E
8-11
8. BRAIN TERMINAL BT200 OPERATION
a. Display Selection (D20: DISP SELECT)
Follow the instructions given to the below to change
the range of integral indication scales.
When USER SET is selected, the user set values of
integral indication and A11: ENGR. OUTPUT
parameter are indicated.
• Example: Set the integral indicator scale to
engineering units display.
The “%” disappears from the
integral indicator display.
ESC
SET
D20:DISP SELECT
USER SET
FEED
NO
OK
SET
D20:DISP SELECT
NORMAL %
Use the or key
to select “USER SET.”
Press the key twice to
enter the setting.
Press the (OK) key.
F0819.EPS
b. Setting User-set Engineering Unit
(D21: DISP UNIT)
This parameter allows entry of the engineering units
to be displayed on the BT200. When the instrument
is shipped, this is set as specified in the order.
Follow the procedure below to change this setting.
This parameter need not be set for % display.
• Example: Set an engineering unit M.
CODE
CAPS
CLR
ESC
SET
D21:DISP UNIT
M
FEED
NO
OK
SET
D21:DISP UNIT
M_
Set “M.”
Press the key twice to
enter the setting.
Press the (OK) key.
F0820.EPS
c. Lower and Higher Range Value Setup
in Engineering Unit (D22: DISP LRV,
D23: DISP HRV)
These parameter items are used to set the lower and
higher range values for the engineering unit display.
When the instrument is shipped, these are set as
specified in the order. Follow the procedure below to
change these settings. Note that these parameters
need not be set for % display.
• Example: Set lower range value (LRV) to –50
and higher range value (HRV) to 50.
FEED
NO
OK
SET
D23:DISP HRV
50M
Press the (OK) key.
F0821.EPS
DEL
CLR
ESC
DATA
DIAG
PRNT
ESC
PARAM
D21:DISP UNT
M
D22:DISP LRV
– 50M
D23:DISP HRV
50M
DEL
CLR
ESC
SET
D22:DISP LRV
0M
- 50
Set “–50.”
Press the key twice to
enter the setting.
Setting LRV
Setting HRV
Set “50.”
Press the key twice to
enter the setting.
SET
D23:DISP HRV
100M
+ 50
(7) Unit Setup for Displayed Temperature
(D30: TEMP UNIT)
When the instrument is shipped, the temperature units
are set to degC. Follow the procedure below to
change this setting. Note that changing the unit here
changes the unit for A20: AMP TEMP (amplifier
temperature) and A21: CAPSULE TEMP (capsule
temperature).
• Example: Change the unit for the temperature
display.
ESC
SET
D30:TEMP UNIT
deg C
< deg C >
< deg F >
Use the or key to
select “deg F.”
Press the key twice to
enter the setting.
F0822.EPS