6) integral indicator scale setup, 6) integral indicator scale setup -10, User-set engineering unit display – Yokogawa EJA530A User Manual
Page 41: For % display, set this parameter only
IM 01C21F01-01E
8-10
8. BRAIN TERMINAL BT200 OPERATION
(6) Integral Indicator Scale Setup
The following 5 displays are available for integral indicators.
T0808.EPS
D20: DISP SELECT
Display
Related Parameters
NORMAL %
USER SET
USER & %
INP PRES
PRES & %
Indicates –5 to 110%
range depending on the
Measurement range
(C21, C22).
Indicates values depending
on the Engineering range
(D22, D23).
(Note 1)
Units set using Engineering
unit (D21) are not indicated.
Indicates user set and %
alternately in 3 second
intervals.
Indicates input
pressure.
(Note 2)
Indication limits –19999
to 19999.
Indicates input pressure
and % alternately in 3
second intervals.
A10:OUTPUT (%)
45.6 %
Description
A11:ENGR.OUTPUT
20.0 M
A10:OUTPUT (%)
45.6 %
A11:ENGR. OUTPUT
20.0 M
A40:INPUT
456 kPa
A10:OUTPUT (%)
45.6 %
A40:INPUT
456 kPa
(Note 1) Scale range can be specified with range limit specifications up to 5 digits (excluding any
decimal point) for low or high range limits within the range of –19999 to 19999.
The range with decimals is available to the third decimal place.
(Note 2) It indicates the value after zeroing.
See (a.) through (c.) for each setting procedure.
% indication and input
pressure indication
Transmitter is set
for “% display” when
shipped.
D20: DISP SELECT
NORMAL %
INP PRES
PRES & %
User-set engineering unit display
D20: DISP SELECT
USER SET
USER & %
D21: DISP UNIT
D22: DISP LRV
D23: DISP HRV
Set for user-set
engineering unit display.
Set a unit to be
displayed on the BT200.
For % display, set this
parameter only.
Set a numeric value for
engineering unit for 4 mA
output (LRV).
Set a numeric value for
engineering unit for 20 mA
output (HRV).
F0818.EPS