3 setting parameters, 1) tag no. setup (c10: tag no) set, 2) calibration range setup – Yokogawa EJA530A User Manual
Page 38: Setting parameters -7, 1) tag no. setup -7 (2) calibration range setup -7, Example: set a tag no. to fic-1a, Example: change the unit from mmh, O to kpa
IM 01C21F01-01E
8-7
8. BRAIN TERMINAL BT200 OPERATION
8.3.3 Setting Parameters
Set or change the parameters as necessary. After
completing these, do not fail to use the “DIAG” key
to confirm that “GOOD” is displayed for the self-
diagnostic result at _60: SELF CHECK.
(1) Tag No. Setup (C10: TAG NO)
Use the procedure below to change the Tag No. Up
to 16 alphanumeric characters can be entered.
Press the key to turn on
the BT200.
Connect DPharp and BT200 using
a communication cable and press
the key.
PARAM
01:MODEL
EJA110A-DM
02:TAG NO.
YOKOGAWA
03:SELF CHECK
GOOD
OK
MENU
A:DISPLAY
B:SENSOR TYPE
HOME
SET
ADJ
ESC
MENU
C:SETTING
D:AUX SET 1
E:AUX SET 2
H:AUTO SET
HOME
SET
ADJ
ESC
MENU
C10:TAG NO.
YOKOGAWA
C20:PRESS UNIT
kPa
C21:LOW RANGE
0 kPa
DATA
DIAG
PRNT
ESC
Select C: SETTING and press the
key.
Select C10: TAG NO. and press
the key.
Displays the name of connected
DPharp model, TAG NO. and
diagnostics information. Press the
(OK) key after confirmation.
Press the (SET) key to
display the SET menu panel.
––WELCOME––
BRAIN TERMINAL
ID: BT200
check connection
push ENTER key
UTIL
FEED
F0810.EPS
SET
C10:TAG NO.
YOKOGAWA
YOKOGAWA
CODE
CAPS
CLR
ESC
When you have made an entry mistake, return
the cursor using the key, then reenter.
Set the new TAG NO. (FIC-1a).
FOKOGAWA
FIKOGAWA
FICOGAWA
FIC-GAWA
FIC-1AWA
FIC-1aWA
FIC-1a
Set TAG NO. and press the
key.
SET
C10:TAG NO.
YOKOGAWA
FIC-1a _
CODE
caps
CLE
ESC
• Example: Set a Tag No. to FIC-1a
SET
C10:TAG NO.
FIC-1a
FEED
NO
OK
DATA
DIAG
PRNT
ESC
This is the panel for confirming
set data. The set data items flash.
When all items have been confir-
med, press the
again. (To go back to the setting
panel, press the (NO) key.
SET
C10:TAG NO.
YOKOGAWA
FIC-1a
PRINTER OFF
F2:PRINTER ON
PARAM
C10:TAG NO.
FIC-1a
C20:PRESS UNIT
kPa
C21:LOW RANGE
0 kPa
FEED
POFF
NO
The DPharp TAG NO. was
overwritten.
Press the (OK) key to
return to the parameter panel.
Press the (NO) key to
return to the setting panel.
F0811.EPS
(2) Calibration Range Setup
a. Setting Calibration Range Unit
(C20: PRESS UNIT)
The unit is set at the factory before shipment if
specified at the time of order. Follow the procedure
below to change the unit.
mmH
2
O
mmAq
mmWG
mmHg
Torr
Pa
hPa
kPa
MPa
mbar
bar
gf/cm
2
kgf/cm
2
inH
2
O
inHg
ftH
2
O
psi
atm
• Example: Change the unit from mmH
2
O to kPa.
ESC
SET
C20:PRESS UNIT
kPa
FEED
NO
OK
SET
C20:PRESS UNIT
mmH20
< mmWG >
< mmHG >
< Torr >
< kPa >
Use the or
key to select “kPa.”
Press the key twice
to enter the setting.
Press the (OK) key.
F0812.EPS