Field calibration – Watlow Series 1500 User Manual

Page 79

background image

Technical Reference

Field Calibration, Appendix

WATLOW Series 1500 User's Manual

79

10. Repeat Steps 5-8 until all the readings are correct with no further adjustments

necessary.

11. Turn power OFF. Set DIP Switch #7 ON (deg. C). Move DVM (+) plus lead to

Signal conditioner connector terminal 11 (terminal 4 for CH-2). Turn power back
ON.

12. Set mV Source as follows, and adjust "Deg. C Lo Pot" for the corresponding

DVM reading. Actual should read the value shown.

Type

mV Source

DVM Reading

Actual Reading

J

-0.885mV, (0

°

F)

-0.028VDC

±

0.003

- 18

°

C

±

3

K

-0.692mV, (0

°

F)

-0.028VDC

±

0.003

- 18

°

C

±

3

R

2.282mV, (550

°

F)

+0.448VDC

±

0.003

287

°

C

±

3

13. Set mV Source as follows, and adjust "Deg. C Hi Pot" for the corresponding

DVM reading. Actual should read the value shown.

Type

mV Source

DVM Reading

Actual Reading

J

39.40mV, (1300

°

F)

1.100VDC

±

0.003

704

°

C 3

K

52.93mV, (2400

°

F)

2.055VDC

±

0.003

1315

°

C

±

3

R

20.28mV, (3100

°

F)

2.662VDC

±

0.003

1704

°

C

±

3

14. Set mV Source as follows. DVM and Actual Readings should be as shown.

Type

mV Source

DVM Reading

Actual Reading

J

17.19mV, (600

°

F)

0.4922VDC

±

0.003

315

°

C

±

3

K

26.97mV, (1200

°

F)

1.014VDC

±

0.003

649

°

C

±

3

R

10.27mV, (1800

°

F)

1.534VDC

±

0.003

982

°

C

±

3

15. Repeat Steps 12-14 until all the readings are correct with no further adjustments

necessary.

16. If calibrating one channel at a time, power-down the Series 1500 and mV Source.

Move the C/J Compensator from the CH-1 connector terminals to the CH-2
terminals, and move the shorting wiring to the CH-1 terminals. Then perform
Steps 6-14 for CH-2.

Steps 10-14 only
required if the deg.
C Analog Output
is used.

NOTE

Field Calibration