beautypg.com

Ma process input, Ma process input procedure – Watlow Calibrating Watlow Series 988 Family Process Controls User Manual

Page 11

background image

WATLOW Series 988 Calibration Manual

11

mA Process Input

mA Process Input Procedure

0 to 20mA and 4 to 20mA units

Equipment Required

Precision current source, 0-20mA range with 0.01mA resolution

Setup and Calibration
1. Connect the AC voltage L1, L2 and ground to the proper terminals on the

Series 988. See Chapter 2 in the user's manual.

2. For Input #1 calibration: Connect the voltage source to terminal #8 (-) and

terminal #10 (+) on the 988 terminal strip. Use 20 to 24 gauge copper wire.
Remove the control from the case and set the Input 1 DIP switches to:
1 ON

2 ON

3 ON

For Input #2 calibration: Connect the voltage source to terminal #18 (-) and
terminal #20 (+) on the 988 terminal strip. Use 20 to 24 gauge copper wire.
Remove the control from the case and set the Input 2 DIP switches to:
1 ON

2 ON

3 ON

3. Apply power to the unit and let it warm up for 15 minutes. After warm up put

the unit in the CAL menu. See Page 6.

When performing calibration procedures, always do the Input 1 prompts
(beginning with an “A”) first, then continue on to the Input 2 prompts
(beginning with a “B”). See the Calibration menu on Page 6.

4. At the A20A or B20A prompt, enter 20.00mA from the current source to the

Series 988. Allow at least 10 seconds to stabilize. Press the Up key to change
the upper display to YES. Press the Mode key.

5. At the A 4A or B 4A prompt, enter 4.00mA from the current source to the

Series 988. Allow at least 10 seconds to stabilize. Press the Up key to change
the upper display to YES. Press the Mode key. To conclude the current
process calibration, press the Display key.

NOTE:

Any prompt begin-
ning with an “A”
applies to Input 1.
Those beginning
with a “B” apply to
Input 2.