To prepare the installation site
Follow these steps to prepare the site for system installation.
C A U T I O N
This product is designed for indoor use in Installation Category II
and Pollution Degree 2 per IEC 61010-1 and 664 respectively.
To prepare the installation site
Step
Notes
1 Ensure that your installation site meets the
environmental requirements.
•
Environmental requirements (temperature,
relative humidity, altitude, and clearance) are
listed in
2 Ensure that your installation site meets the
electrical requirements.
•
Electrical requirements are listed in
Table 16 Environmental Requirements
Environmental Parameter
System Requirement
Temperature
+0
°
C to +45
°
C (+32
°
F to +113
°
F)
* †
* Install air conditioning and heating as needed to achieve the required ambient temperature range
.
† Accuracy-enhanced measurement performance is specified at an ambient temperature range of
+25
°
C ±5
°
C. After calibration, hold the ambient temperature of the measurement environment to
±1
°
C of the ambient temperature at the time of calibration.
Relative humidity
Maximum 80% for temperatures up to 31
°
C
decreasing linearly to 50% at 40
°
C
Altitude
Up to 3000 meters (approximately 10000 feet)
Clearance (behind and above rack)
‡
‡ Required to ensure the extractor fans can properly ventilate the system.
15 centimeters (6 inches) minimum
Table 17 Electrical Requirements
Electrical Parameter
System Requirement
Supply capability
100/120 volts, 2000 VA
200/240 volts, 2000 VA
Circuit sharing
Do not connect air conditioning or
motor-operated equipment to the same ac
circuit supplying line voltage to the system.