Alignment Procedures

97

Alignment Program (continued)

1050 !
1060 !
1070 !
1080 DEF FNDci(I_range)

! Function to read current

1090 COM

/Vm/

@Vm

1100

COM /Shunt/ Shunt_r

1110

IF I_range=2 THEN

! 2A must read across shunt

1120

Amps=FNDcv/Shunt_r

! Convert shunt voltage to Amps

1130

ELSE!

! <2A read direct from DMM

1140

OUTPUT @Vm;"DCI"

! Set DMM to read DC current

1150

TRIGGER @Vm

! Take a current reading

1160

ENTER @Vm;Amps

! Enter the current

1170 END

IF

1180

RETURN Amps

! Return the current reading

1190 FNEND
1200 !
1210 !
1220 !
1230 SUB Output_on

! Subprogram to turn ON outputs

1240

COM /Ps/ Chan,Other_chan, @Ps, Model$[7]

1250

COM /Outputs/ No_of_outputs

1260

OUTPUT @Ps; OUT l,l;OUT 2,1

! Turn on outputs 1 & 2

1270

IF No_of_outputs=4 THEN OUTPUT @Ps;"OUT 3,1;OUT 4,1"

1280

! Turn on outputs 3 & 4

1290 SUBEND
1300 !
1310 !
1320 !
1330 SUB Output_off

! Subprogram to turn OFF outputs

1340

COM /Ps / Chan, Other_chan, @Ps, Mode1$ [ 7 ]

1350

COM /Outputs/ No_of_outputs

1360

OUTPUT @Ps;"OUT 1,0,OUT 2,0"

! Turn off outputs 1 & 2

1370

IF No_of_outputs=4 THEN OUTPUT @Ps; “OUT 3,0;OUT 4,0”

1380

! Turn off outputs 3 & 4

1390 SUBEND
1400 !
1410 !
1420 !
1430 Cal_voltage : SUB Cal_voltage(V_range)

! Subprogram to cal voltage

1440

COM /Ps/ Chan,Other_chan,@Ps,Model$ [7]

1450 !
1460

IF V_range=50 THEN GOTO Cal_v

! Skip setup instructions

1470

Output_off

! Turn off all outputs

1480

Clear_screen

1490

PRINT "Set up output";Chan;"for voltage calibration

1500

PRINT

1510

PRINT " Use Fig A-1 of the Operating Manual"

1520

PRINT

1530

PRINT “ 3458A input is connected to VOLTAGE terminals”

1540

PRINT

1550

PRINT " The current shunt is NOT used"

1560

PRINT