Numeric data – Agilent Technologies 6625A User Manual
Page 67
Remote Operation
67
Table 5-1. Power Supply Commands (continued)
Command
Header
Output
Channel
Data Range
(Fig. 5-2)
Syntax
Store present output state
STO
0-10
C2
Query preset status of output
STS?
1, 2, 3, 4
Q2
Perform self test on GP-IB interface
TEST?
.
C1
Set bits in mask register
UNMASK
1, 2, 3, 4
0-255
C4
Query bits set in mask register
UNMASK?
1, 2, 3, 4
Q2
Program the voltage DAC in counts
VDAC
1, 2, 3, 4
See Service Manual
C4
Querys setting of voltage DAC in counts
VDAC?
1, 2, 3, 4
Q2
Send data to calibrate the voltage circuits
VDATA
1, 2, 3, 4
See Table 5-4
C5
Set output to high V calibration value
VHI
1, 2, 3, 4
See Table 5-4
C3
Set output to low V calibration value
VLO
1, 2, 3, 4
See Table 5-4 V
C3
Query inputs of analog multiplexer
VMUX?
1, 2, 3, 4
1-18
C4
Query measured value of an output
VOUT?
1, 2, 3, 4
See Table 5-4
Q2
Calibrate the voltage readback circuitry
VRDAT
1, 2, 3, 4
See Table 5-4
C5
Set output to V readback high cal value
VRHI
1 2, 3, 4
See Table 5-4
C3
Set output to V readback low cal value
VRLO
1, 2, 3, 4
See Table 5-4
C3
Set full scale voltage programming range
VRSET
1, 2, 3, 4
See Table 5-4
C4
Query full scale voltage programming range
VRSET?
1, 2, 3, 4
See Table 5-4
Q2
Set output voltage
VSET
1, 2, 3, 4
See Table 5-4
Q2
Query setting of output voltage
VSET?
1, 2, 3, 4
See Table 5-4
Q2
Increase or decrease output voltage by value
VSTEP
1, 2, 3, 4
See Table 5-4
C4
Figure 5-2 shows the possible syntax forms for the device commands that are used to program the power supply. Syntax
forms for the calibration commands that are discussed in Appendix A are also included. The oblong shape at the left of the
syntax forms contains the command header which must be entered as shown in Tables 5-1 and 5-2. Commands are accepted
in either uppercase or lowercase letters (ASCII characters). Circles contain characters that must be entered exactly as
shown. Characters such as a space
as a line feed < LF > or a semicolon are used to terminate the command string. Rectangles contain parameters that follow
the command header lines and arrows indicate the correct paths through the syntax diagrams.
Numeric Data
The power supply will accept numeric data in implicit point, explicit point, or scientific notation. A general syntax diagram
for numeric data is included in Figure 5-2. Implicit point notation means that numbers do not contain a decimal point;
integers for example. Numbers written in explicit notation contain a decimal point, such as 12.35. In scientific notation, the
letter E stands for "10 raised to". For example, 1.2E3 is read as 1.2 times 10 raised to the 3rd power, which equals 1,200.
Plus and minus signs are considered numeric characters and are optional. If you program a number with an accuracy that is
greater than the resolution of the supply, the number will automatically be rounded to the nearest multiple of the power
supply’s resolution. Table 5-1 gives the ranges for numeric data that is sent to the supply.
The power supply will also return numeric data (ASCII characters) to your computer. The format of the numbers returned
depends upon the type of data requested. Table 5-2 gives the format for data returned to the computer in response to any of
the queries that are listed.