Alignment Procedures

100

Alignment Program (continued)

2610

Irlo=FNDci(I_range)

! Read low output current

2620

!

2630

OUTPUT @Ps;"IRHI";Chan

! Set high readback current

2640

REPEAT

! Wait for supply to finish

2650

UNTIL BIT(SPOLL(@Ps),4)

! Finished when Bit 4 goes true

2660

Irhi=FNDci(I_range)

! Read high output current

2670

!

2680

OUTPUT @Ps;"IRDAT";Chan,Irlo,Irhi

! Send current readback data

2690

!

2700 SUBEND
2710 !
2720 !
2730 !
2740 Cal_sink SUB Cal_sink(I_range,Polarity)

! Subprogram to cal -CC

2750

COM /Ps/ Chan,Other_chan,@Ps,Model$[7]

2760

!

2770

IF I_range= .5 THEN GOTO Sink_rb

! Skip setup instructions

2780

IF Model$="Agilent6625A" AND I_range=2 THEN I_range= .5

2790

! If a 6625A, 2A output will be

2800

! set to 5A to match output 1

2810

IF Chan=2 OR Chan=4 THEN GOTO Sink_rb

! Skip setup instructions

2820 !
2830 Output_off
2840 Clear_screen
2850 PRINT “Set up output';Chan;”for ;I_range; “A current SINK cal”
2860 PRINT
2870

PRINT “ Connect outputs”;Chan; “and” ;Other_chan; “together as in”

2880 !
2890

IF I range=2 THEN

! If it is 2A output use setup 5

2900

PRINT " Fig A-5 of the Operating Manual'

2910

PRINT

2920

PRINT " 3458A input is connected to VOLTAGE terminals"

2930

PRINT

2940

PRINT " The current shunt IS used for this part"

2950

PRINT " of the alignment"

2960

ELSE

! If not 6626A 2A use setup 4

2970

PRINT “Fig A-4 of the Operating Manual”

2980

PRINT

2990

PRINT “3458A input is connected to CURRENT terminals”

3000

PRINT

3010

PRINT " The current shunt IS NOT used"

3020

END IF

3030

PRINT

3040

PRINT “The sink readback of both outputs”

3050

PRINT “will be aligned with this setup”

3060

!

3070 PRINT
3080

PRINT "Press CONTINUE when ready”

3090

PAUSE

3100

!

3110 Sink_rb
3120 Clear_screen