Agilent Technologies 6625A User Manual

Page 66

background image

Remote Operation

66

Table 5-1. Power Supply Commands (continued)

Command

Header

Output

Channel

Data Range

(Fig. 5-2)

Syntax

Query present hardware error

ERR?

---

See Table 5-8

Q1

Query fault register

FAULT?

1, 2, 3, 4

---

Q2

Query the model number of supply

ID?

---

---

Q1

Program the I DAC in counts

IDAC

1, 2, 3, 4

See Service Manual

C4

Query setting of I DAC in counts

IDAC?

1, 2, 3, 4

---

Q2

Send data to calibrate I circuits

IDATA

1, 2, 3, 4

See Table 5-4

C5

Sets output to high I cal. value

IHI

1, 2, 3, 4

See Table 5-4

C3

Sets output to low I cal. value

ILO

1, 2, 3, 4

See Table 5-4

C3

Query measured I output

IOUT?

1, 2, 3, 4

See Table 5-4

Q2

Used to Calibrate I readback circuits

IRDAT

1, 2, 3, 4

See Table 5-4

C5

Set output to + I readback high cal value

IRHI

1, 2, 3, 4

See Table 5-4

C3

Set output to - I readback high cal value

IRHN

1, 2, 3, 4

See Table 5-4

C3

Set output to - I readback low cal value

IRLN

1, 2, 3, 4

See Table 5-4

C3

Set output to + I readback low cal value

IRLO

1, 2, 3, 4

See Table 5-4

C3

Set full scale current range of output

IRSET

1, 2, 3, 4

See Table 5-4

C4

Query full scale current range

ISET?

1, 2, 3, 4

See Table 5-4

Q2

Set current of an output

ISET

1, 2, 3, 4

See Table 5-4

C4

Query current of an output

ISET?

1, 2, 3, 4

See Table 5-4

Q2

Increase or decrease output current by value

ISTEP

1, 2, 3, 4

See Table 5-4

C4

Select which output will be metered

METER

---

1-4

C2

Query which output is being metered

METER?

---

---

Q1

Send data to calibrate -I readback

NIDAT

1, 2, 3, 4

See Table 5-4

C5

Enable overcurrent protection

OCP

1, 2, 3, 4

0,1 (OFF,ON)

C4

Query if OCP is enabled

OCP?

1, 2, 3, 4

---

Q2

Reset overcurrent protection

OCRST

1, 2, 3, 4

---

C3

Turn output on or off

OUT

1, 2, 3, 4

0,1 (OFF,ON)

C4

Query if output is on or off

OUT?

1, 2, 3, 4

---

Q2

Perform overvoltage calibration

OVCAL

1, 2, 3, 4

See Table 5-4

C3

Reset overvoltage circuit

OVRST

1, 2, 3, 4

---

C3

Set overvoltage trip value

OVSET

1, 2, 3, 4

See Table 5-4

C4

Query overvoltage trip value

OVSET?

1, 2, 3, 4

See Table 5-4

Q2

Enable power on service request

PON

---

0,1 (OFF,ON)

C2

Query if PON is enabled

PON?

1, 2, 3, 4

---

Q1

Recall voltage and current settings

RCL

---

0-10

C2

Set readback DAC to a number of counts

RDAC

1, 2, 3, 4

See Service Manual

C4

Query readback DAC count setting

RDAC?

1, 2, 3, 4

---

Q2

Query revision date of ROM

ROM?

---

Q1

Query revision date of secondary ROM

SROM?

1, 2, 3, 4

---

Q2

Set causes for generating a service request

SRQ

---

01,2,3

C2

Query causes which will generate an SRQ

SRQ?

---

---

Q1