Agilent Technologies 6625A User Manual
Page 17
General Information
17
Table 1-2. Specifications (continued)
Source Effect:
Voltage
0.5 mV
0.5 mV
0.5 mV
0.5 mV
+ Current
0.005 mA
0.005 mA
0.01 mA
0.01 mA
Programming Accuracy:
Note: The programming accuracy specifications may degrade slightly when the unit is subjected to an RF field equal
to or greater than 3 volts/meter.
Voltage
0.016% + 1.5 mV
0.016% + 10 mV
0.016% + 3 mV
0.016% + 10 mV
+ Current
0.04% + 15
µ
A
0.04% + 100
µ
A
0.04% + 185
µ
A
0.04% + 500
µ
A
OVP
0.13% + 475 mV
0.13% + 475 mV
0.13% + 475 mV
0.13% + 475 mV
Measurement/Readback Accuracy*
Voltage
0.016% + 2 mV
0.016% + 10 mV
0.016% + 3.5 mV
0.016% + 10 mV
±
Current
0.03% + 15
µ
A
0.03% + 130
µ
A
0.04% + 250
µ
A
0.04% + 550
µ
A
Load Transient Recovery Time
:(all outputs)
75
µ
S maximum to recover to within 75 mV of nominal value following a load change from 0.1 A to 100% of the
maximum rated current
Maximum Output Noise:
(PARD) (20 Hz-20 MHz)
CV pk-to-pk
3 mV
3 mV
3 mV
3 mV
CV rms
0.5 mV
0.5 mV
0.5 mV
0.5 mV
+CC rms
0.1 mA
0.1 mA
0.1 mA
0.1 mA
AC Input Voltage and Frequency:
Nominal Line l00, 120, 220, or 240 Vac
Amplitude +6% to -13% of nominal line voltage
Frequency 47 to 66 Hz
Note at low line, the supply will operate with up to 3/4 ohm line resistance
*For a
±
5 ° C range about the calibration temperature
Table 1-3. Supplemental Characteristics
Output Power
25 Watt Output
50 Watt Output
Output Range
Lo Range
Hi Range
Lo Range
Hi Range
Output Volts
0-7 V
0-50 V
0-16 V
0-50 V
Output Amps
0-15 mA
0-500 mA
0-200 mA
0-2 A
Temperature Coefficient-Programming:
Voltage
(0.003%+0.1 mV)
per° C
(0.003%+0.1 mV)
per° C
(0.003%+0.1 mV)
per° C
(0.003%+0.1 mV)
per° C
+ Current
(0.0035% + 1.5
µ
A)
per° C
(0.0035% + 6
µ
A)
per° C
(0.0035% + 30
µ
A)
per° C
(0.0035% + 50
µ
A)
per° C
OVP
(0.013% + 1 mV)
per° C
(0.013% + 2 mV)
per° C
(0.013% + 1 mV)
per° C
(0.013% + 2 mV)
per° C