Error Codes and Messages

119

Table D-2. ERROR Responses (continued)

Error Code

(ERR? query)

Message

(ERR key)

Explanation

18

CAL LOCKED

Calibration was attempted with the calibration jumper on the GP-IB
board in the lockout position (See Section IV in the Service Manual).
Reposition the jumper and re-calibrate if this is desired.

22

SKIP SLF TST

The self test jumper on the GP-IB board is in the Skip Self Test Position
(See Section IV in the Service Manual). Reposition the jumper and carry
out self-test if this is desired.

Table D3. TEST? Response

Response Code

(TEST? query)

Explanation

0

No errors were detected.

20

The timer on the GP-IB board failed self test. Refer to the Troubleshooting Section in the
Service Manual.

21

The RAM on the GP-IB Board failed self test. Refer to the Troubleshooting Section in the
Service Manual. Note that if this failure occurred at power-on, it cannot be displayed on the
front panel.

27

The RAM on the GP-IB Board failed self test. Refer to the Troubleshooting Section in the
Service Manual. Note that if this failure occurred at power-on, it cannot be displayed on the
front panel.

29

See Table 5-9 on page 84.

30

See Table 5-9 on page 84.