LTCA-10/LTCA-40 USER’S MANUAL

31

h. The “REFERENCE TEMP” screen will be displayed:

Use the

[PAPER Contrast]

and

[PAPER Contrast]

keys to adjust the

Reference Temperature. Press the

[ENTER]

key to confirm the temperature selection.

NOTE

The LTCA-10/40 will calculate the equivalent resistance value at this new
temperature.

i. The following “WARNING” screen will be displayed:

This warning reminds the operator that the next sequence of test steps will run current
through the test load. Press the

[ENTER]

key to proceed to the next step.

j. One of the following menus will be displayed (with the relevant test current value)

depending on how many channels were selected in step c.

V1 Test Only

V1 & V2 Test

V1, V2, & V3 Test

If you have made any errors or would like to abort the test, press the

[STOP]

key, and

you will be returned to the “START-UP” menu.

-V1, V2, & V3 TEST-

1 AMP TEST

“START” TO RUN TEST

OR

“STOP” TO ABORT

-V1 & V2 TEST-

1 AMP TEST

“START” TO RUN TEST

OR

“STOP” TO ABORT

-V1 TEST ONLY-

1 AMP TEST

“START” TO RUN TEST

OR

“STOP” TO ABORT

*********WARNING*********

DANGEROUS FLASH-OVER
WILL OCCUR IF CABLES

ARE DISCONNECTED!

*************************

REFERENCE TEMP:

85.0

°C 185.0°F


↑↓ to adjust temp

“ENTER” to accept