LTCA-10/LTCA-40 USER’S MANUAL

21

a. Press the

[1]

key to select copper as the winding material. Continue to

step f.

b. Press the

[2]

key to select aluminum as the winding material. Continue to

step f.

c. Press the

[3]

key to manually enter the Tk value. The following screen

will be displayed:

You can increase the Tk value by 0.5

°C increments by pressing the

[PAPER Contrast]

key. You can decrease the Tk value by 0.5

°C

increments by pressing the

[PAPER Contrast]

key. Press the

[ENTER]

key to confirm your temperature selection. Continue to step f.

2. NO

If you do NOT wish to convert readings to standard temperature, press the

[2]

key (NO) and continue to step h.

f. After the Tk value has been selected, the “D.U.T. TEMP” screen will be displayed:

Use the

[PAPER Contrast]

and

[PAPER Contrast]

keys to adjust the D.U.T.

temperature. Press the

[ENTER]

key to confirm the temperature selection.

NOTE

D.U.T is short for Device Under Test

g. The “REFERENCE TEMP” screen will be displayed:

Use the

[PAPER Contrast]

and

[PAPER Contrast]

keys to adjust the

Reference Temperature. Press the

[ENTER]

key to confirm the temperature selection.

REFERENCE TEMP:

85.0

°C 185.0°F


↑↓ to adjust temp

“ENTER” to accept

D.U.T. TEMP

25.0

°C 77.0°F


↑↓ to adjust temp

“ENTER” to accept

Tk:

230.0

°C


↑↓ to adjust Tk

“ENTER” to accept