LTCA-10/LTCA-40 USER’S MANUAL
21
a. Press the
[1]
key to select copper as the winding material. Continue to
step f.
b. Press the
[2]
key to select aluminum as the winding material. Continue to
step f.
c. Press the
[3]
key to manually enter the Tk value. The following screen
will be displayed:
You can increase the Tk value by 0.5
°C increments by pressing the
[PAPER ∧ Contrast]
key. You can decrease the Tk value by 0.5
°C
increments by pressing the
[PAPER ∨ Contrast]
key. Press the
[ENTER]
key to confirm your temperature selection. Continue to step f.
2. NO
If you do NOT wish to convert readings to standard temperature, press the
[2]
key (NO) and continue to step h.
f. After the Tk value has been selected, the “D.U.T. TEMP” screen will be displayed:
Use the
[PAPER ∧ Contrast]
and
[PAPER ∨ Contrast]
keys to adjust the D.U.T.
temperature. Press the
[ENTER]
key to confirm the temperature selection.
NOTE
D.U.T is short for Device Under Test
g. The “REFERENCE TEMP” screen will be displayed:
Use the
[PAPER ∧ Contrast]
and
[PAPER ∨ Contrast]
keys to adjust the
Reference Temperature. Press the
[ENTER]
key to confirm the temperature selection.
REFERENCE TEMP:
85.0
°C 185.0°F
↑↓ to adjust temp
“ENTER” to accept
D.U.T. TEMP
25.0
°C 77.0°F
↑↓ to adjust temp
“ENTER” to accept
Tk:
230.0
°C
↑↓ to adjust Tk
“ENTER” to accept