Vanguard EZCT-2000 User Manual
Page 32

EZCT-2000 VERSION 1.x SOFTWARE MANUAL
REV 1
30
5.0 WORKING WITH TEST PLANS
The EZCT-2000 software can be used to create current-transformer test plans. Test plans can
then be run from the PC or transferred to the CT Tester to be run from the CT Tester (in stand-
alone mode). Test plans can also be retrieved from a CT Tester using the EZCT-2000 software.
5.1
Retrieving Test Plans From a CT Tester
To retrieve a test plan from a CT Tester:
1. Make sure the EZCT-2000 software is running. Connect the CT Tester to the PC via either
the RS-232C port or the USB port.
2. From the Test-Plan menu select Retrieve Test Plan from device (alternatively you can
click on the File menu, click on Test Plan and then select Retrieve Test Plan from device).
A window will appear listing a directory of all the test plans stored in the CT Tester’s
memory as shown below:
3. You can select a test plan to be retrieved by clicking on the test plan number. The
selected test plan will be highlighted. You may select multiple test plans by holding
down the [CTRL] key and clicking on the test plan numbers. All selected test plans will
be highlighted. You may de-select a selected test plan by holding down the [CTRL] key
and clicking on the selected test plan number a second time. To select all of the test
plans, click on the “Select All” button.
4. The “Filename Prefix” input field allows you to enter a word that will be used as part of
the filename for the stored test plan on the PC hard drive. When a test plan is retrieved
from a CT Tester and stored on the hard drive, the filename is in the following format: