ATO-400 / ATO-600 Operating Procedures

36

6.11 Calibration Test Procedure

This procedure describes the steps to verify the ATO mesurement accuracy and generates a
calibration report. Five different calibration shunts are required for this test: 50micro-ohm, 100
micro-ohm, 500 micro-ohm, 1000 micro-ohm, 1900 micro-ohm. These test shunts shall be 0.1%
accuracy or better.

Table 12.0 Calibration Test Procedure

STEP

ACTION

ATO DISPLAY

12-1

Press key # 2 on the START MENU to go
to the SETUP MENU shown at right.

12-2

On the SETUP MENU, press key #4
(NEXT PAGE).

12-3

Press key #2 to select “CALIBRATE
UNIT”.

12-4

Press “ENTER” key to continue.

12-5

Connect the 50 micro-ohm shunt to ATO.
Press “ENTER” key to start test.

12-6

The ATO will measure the 50

?

-ohm shunt

using 10A test current.
Observe test results on LCD.

12-7

Press “ENTER” key to continue.
The ATO will measure 50

?

-ohm shunt

using 100A test current.

12-8

Observe test results. Press “ENTER” key to
continue.

1. ENTER RECORD ID
2. REVIEW RECORD
3. RESTORE RECORD
4. NEXT PAGE

1. SET TIME
2. CALIBRATE UNIT

CALIBRATION-SPECIFIC
SHUNTS ARE REQUIRED!
“ENTER” TO CONTINUE

CONNECT 50

? -OHM

0.1% SHUNT NOW… .
“ENTER” TO CONTINUE

CURRENT: 10.0 AMP
BURN IN: 02 SEC
50

? -OHM TEST

FINAL RESISTANCE
I= 9.9 AMPS
50.5 MICRO-OHM
ANY KEY TO CONTINUE

CURRENT: 100.0 AMP
BURN IN: 02 SEC
50

? -OHM TEST

FINAL RESISTANCE
I= 100.0 AMPS
50.4 MICRO-OHM
ANY KEY TO CONTINUE