beautypg.com

Vanguard ATO-600 User Manual

Page 23

background image

ATO-400 / ATO-600 Operating Procedures

23

Table 5.0 Run Test Procedure (Measure an Unknown Resistance Continued)

STEP

ACTION

ATO DISPLAY

5-10

“RUN ANOTHER TEST?” displays.

5-11

If another test is required, press key #1
(YES). The SELECT TEST CURRENT
menu will appear.Go to step 5-12.
If another test is not required, then press
key #2 (NO) and go to step 5-15.

5-12

Return to step 5-2 to setup for another test
Repeat the sequence from step 5-2 to step
5-10.

Step 5-2 thru 5-10 displays repeat.

5-13

“SAVE THIS RECORD?” displays when
another test was not selected in step 5-12.
The test record contains all the readings
taken in this test.

5-14

To save this record, press key #1 (YES). If
the test record is not to be saved, press key
#2 (NO) and go to step 5-16.

NOTE

A test record is saved in Flash
EEPROM. A record number
will be automatically assigned
to the record by the ATO.
(The Test record was assigned
#2 in this example).

5-15

When the record is saved the RECORD
NUMBER “x” shows on display. Press
ENTER to return to the START MENU.

5-16

Pressing key #2 (NO) on SAVE THIS
TEST RECORD (see step 5-14) causes the
ARE YOU SURE? prompt (shown at
right).

ARE YOU SURE?
DATA WILL BE LOST!
1. DO NOT SAVE RECORD
2. SAVE RECORD

1. RUN TEST 04/11/05
2. SETUP 12:24:01
3. DIAGNOSTIC

RECORD NUMBER 2
HAS BEEN SAVED!

SAVE THIS TEST RECORD?
1. YES
2. NO

SELECT TEST CURRENT:
1. 10A 2. 25A 3. 50A
4. 100A 5. 200A 6. 300A
7. 400A 8. CUSTOM

RUN ANOTHER TEST?
1. YES
2. NO

This manual is related to the following products: