beautypg.com

Ht/d installation instructions, 4 maintenance, 5 disposal – TREND HT_D User Manual

Page 4: Trend control systems limited, Trend control systems usa, Weee directive, Replace filter, Remove lid, Preparation, Locate buttons

background image

4

HT/D Duct Humidity and Temperature Sensors Installation Instructions TG200988 Issue 1/A 30/01/08

HT/D

Installation Instructions

Please send any comments on this or any other Trend technical publication to [email protected]

Manufactured for and on behalf of the Environmental and Combustion Controls Division of Honeywell Technologies Sàrl, Ecublens, Route du

Bois 37,Switzerland by its Authorized Representative, Trend Control Systems Limited.

©Trend Control Systems Limited 2007. Trend Control Systems Limited reserves the right to revise this publication from time to time and make

changes to the content hereof without obligation to notify any person of such revisions or changes.

Trend Control Systems Limited

P.O. Box 34, Horsham, West Sussex, RH12 2YF, UK. Tel:+44 (0)1403 211888 Fax:+44 (0)1403 241608 www.trend-controls.com

Trend Control Systems USA

6670 185th Avenue NE, Redmond, Washington 98052, USA. Tel: (425)897-3900, Fax: (425)869-8445 www.trend-controls.com

D O W

N

U P

WEEE Directive :

At the end of their useful life the packaging

and product should be disposed of by a

suitable recycling centre.

Do not dispose of with normal household waste. Do not burn.

4 MAINTENANCE

Over time, the sensing element may become covered in dust. The dust can be removed using compressed air. Under no

circumstances should water or cleansing agents be used on the sensing elements.

It is recommended that the accuracy of the sensor is verified every 12 months. If the sensor falls outside the quoted accuracy,

replace the filter and recalibrate as shown below

5 DISPOSAL

Replace filter

1

b

a

3

Remove lid

2

Preparation

Keep sensor and humidity
chamber in same room for 4
hours before.

Place probe in humidity chamber
30 mins before.

Keep temperature constant
during calibration.

4

Locate buttons

push
buttons

calibration
LED

5

Two point calibration

For calibration over whole working range

Start calibration at lower humidity point.

Difference between the two points should be >30%RH

Low Humidity Calibration

1

Remove lid from the calibration chamber (ACC/CAL/HT) and

clean chamber thoroughly.

2

Insert one of the cloths provided with the calibrating liquid. Pour

calibrating liquid (ACC/CAL/HT/35%) onto the fabric. Refit lid

and screw tightly.

3

Insert probe into sensor aperture of chamber and tighten collar to

provide air tight seal around the probe. Wait 30 min.

4

Press Down pushbutton for 3 s to start. LED will illuminate.

5

Press Up or Down pushbutton to adjust measured value in 0.1%

steps.

6

Either: Press Up pushbutton for 3 s to stop. Calibrated value

is stored. LED is extinguished.
Or: Press Down pushbutton for 3 s to exit calibration without

storing value. LED is extinguished.

High Humidity calibration

7

Remove lid from the calibration chamber (ACC/CAL/HT) and

clean chamber thoroughly.

8

Insert one of the cloths provided with the calibrating liquid. Pour

calibrating liquid (ACC/CAL/HT/80%) onto the fabric. Refit lid

and screw tightly.

9

Insert probe into sensor aperture of chamber and tighten collar

to provide air tight seal around the probe. Wait 30 min.

10 Press Up pushbutton for 3 secs to start. LED will illuminate.

11 Press Up or Down pushbutton to adjust measured value in 0.1% steps
12 Either: Press Up pushbutton for 3 s to stop. Calibrated value

is stored. LED is extinguished.
Or: Press Down pushbutton for 3 s to exit calibration without

storing value. LED is extinguished.

6

One point calibration

For calibration over limited range about single point

This calibration decreases accuracy over remainder of working range

Single Point Humidity Calibration

1

Remove lid from the calibration chamber (ACC/CAL/HT) and

clean chamber thoroughly.

2

Insert one of the cloths provided with the calibrating liquid. Pour

calibrating liquid (ACC/CAL/HT/35% or /80%) onto the fabric.

Refit lid and screw tightly.

3

Insert probe into sensor aperture of chamber and tighten collar to

provide air tight seal around the probe. Wait 30 min.

2

Either: (If chamber humidity >50%RH.) Press Up pushbutton

for 3 s to start. LED will illuminate.
Or: (If chamber humidity <50%RH.) Press Down pushbutton for

3 s to start. LED will illuminate.

3

Press Up or Down pushbutton to adjust measured value in 0.1%

steps

4

Either: Press Up pushbutton for 3 s to stop. Calibrated value

is stored. LED is extinguished.
Or: Press Down pushbutton for 3 s to exit calibration without

storing value. LED is extinguished.

Caution: This unit contains static sensitive

devices. Suitable anti-static

precautions should be taken

throughtout the operation to prevent

damage to the units.

BS EN100015/1 Basic Specification: protection

of electrostatic sensitive devices.

Replace Lid

7

Test System

Δ T

Δ H

IQ

8

b

a

unscrew

c

ACC/HTD/FILTER

d

Caution: Do not touch humidity sensing element

See Section 5,
Disposal