beautypg.com

Ht/o installation instructions 4 maintenance – TREND HT_O User Manual

Page 6

background image

6

HT/O Installation Instructions TG200992 Issue 1/A 30/01/08

HT/O

Installation Instructions

4 Maintenance

(continued)

8

One point calibration

For calibration over limited range about single point

This calibration decreases accuracy over remainder of
working range

Single Point Humidity Calibration
1

Remove lid from the calibration chamber (ACC/CAL/HT)
and clean chamber thoroughly.

2

Insert one of the cloths provided with the calibrating
liquid. Pour calibrating liquid (ACC/CAL/HT/35% or /
80%) onto the fabric. Refit lid and screw tightly.

3

Insert probe into sensor aperture of chamber and tighten
collar to provide air tight seal around the probe. Wait
30 min.

4

Either: (If chamber humidity >50%RH.) Press Up
pushbutton for 3 s to start. LED will illuminate.
Or: (If chamber humidity <50%RH.) Press Down
pushbutton for 3 s to start. LED will illuminate.

5

Press Up or Down pushbutton to adjust measured value
in 0.1% steps

6

Either: Press Up pushbutton for 3 s to stop. Calibrated
value is stored. LED is extinguished.
Or: Press Down pushbutton for 3 s to exit calibration
without storing value. LED is extinguished.

Replace Cover

11

Re-attach Sensor

10

Replace Radiation Shield

9

7

Two point calibration

For calibration over whole working range

Start calibration at lower humidity point.

Difference between the two points should be >30%RH

Low Humidity Calibration
1

Remove lid from the calibration chamber (ACC/CAL/HT)
and clean chamber thoroughly.

2

Insert one of the cloths provided with the calibrating
liquid. Pour calibrating liquid (ACC/CAL/HT/35%) onto
the fabric. Refit lid and screw tightly.

3

Insert probe into sensor aperture of chamber and tighten
collar to provide air tight seal around the probe. Wait
30 min.

4

Press Down pushbutton for 3 s to start. LED will illuminate.

5

Press Up or Down pushbutton to adjust measured value
in 0.1% steps.

6

Either: Press Up pushbutton for 3 s to stop. Calibrated
value is stored. LED is extinguished.
Or: Press Down pushbutton for 3 s to exit calibration
without storing value. LED is extinguished.

High Humidity calibration
7

Remove lid from the calibration chamber (ACC/CAL/HT)
and clean chamber thoroughly.

8

Insert one of the cloths provided with the calibrating
liquid. Pour calibrating liquid (ACC/CAL/HT/80%) onto
the fabric. Refit lid and screw tightly.

9

Insert probe into sensor aperture of chamber and tighten
collar to provide air tight seal around the probe. Wait
30 min.

10 Press Up pushbutton for 3 s to start. LED will illuminate.
11 Press Up or Down pushbutton to adjust measured value in

0.1% steps

12 Either: Press Up pushbutton for 3 s to stop. Calibrated

value is stored. LED is extinguished.
Or: Press Down pushbutton for 3 s to exit calibration
without storing value. LED is extinguished.