beautypg.com

Customizing qualiphy, Copy configuration, Select tests – Teledyne LeCroy QPHY-PCIe3-Tx-Rx User Manual

Page 13

background image

QPHY-PCIE3-TxRx

Operator’s Manual

923662 Rev B

9

Customizing QualiPHY

The pre-loaded configurations cannot be modified. However, you can create your own test
configurations by copying one of the pre-loaded configurations and modifying it.

Copy Configuration

1. Access the QualiPHY framework dialog and select a Standard.

2. Click Edit/View Configuration and select the configuration upon which to base the new

configuration. This can be a pre-loaded configuration or another copy.

3. Click Copy and enter a name and description. Once a custom configuration is defined, it

appears on the Configuration tab followed by “(Copy).”

4. Select the new, custom configuration and follow the procedures below to continue

making changes.

Note: If any part of a configuration is changed, the Save As button becomes active on
the bottom of the dialog. If a custom configuration is changed, the Save button will also
become active to apply the changes to the existing configuration.

Select Tests

On the Test Selector tab, check the tests that make up the configuration. Each test is defined
by the PCI CEM standard. A description of each test is displayed when it is selected.

To loop any of the tests in this configuration, select the test from the list, then choose to loop
indefinitely until stopped or enter the number of repetitions.