Decoding pump mode fields – SKC DatraTrac 2000 for the AirChek 2000_3000 User Manual
Page 44
42
Decoding Pump Mode Fields
The Pump Mode fi eld must be converted to hexadecimal numbers. Many scientifi c
calculators are capable of this conversion. For example:
Decimal
Hex
257
101
1314
522
1282
502
258
102
1
001
1025
401
261
105
262
106
Use the following values (or mask bits) for the hexadecimal number to determine
the pump mode:
HOLD
001
RUN
002
FAULT
004
SLEEP
008
LOW BATTERY 010
SET FLOW
020
PROGHOLD 040
PROGRUN
080
STEL
100
LOCKOUT
200
USER SETUP
400
RESET
800
SINGLE POINT
CAL
1000
Then add the bits to determine the code:
101 = STEL + HOLD
522 = USER SETUP + STEL + SET FLOW + RUN
502 = USER SETUP + STEL + RUN
102 = STEL + RUN
001 = HOLD
401 = USER SETUP + HOLD
105 = STEL + FAULT + HOLD
106 = STEL + FAULT + RUN
For example:
Code 1314 (hex 522): pump was in the user interface having its flow (or flow
correction) adjusted.
Code 258 (hex 102): pump was running a STEL.
Transferring History Files