beautypg.com

Setup – SKC 210-5000 AirChek XR5000 User Manual

Page 8

background image

5

Power supply

jack

Charge status

LED

See http://www.skcinc.com/instructions/1756.pdf for more information

on SKC pump battery packs.

Reading the Charging Status LED on the Single Charger

The Li-Ion Charging Unit (Cat. No. P22300) indicates battery charge status via
an LED on the unit that blinks in specifi c patterns. Observe the LED steadily for
> 5 seconds to read charge status.

LED Action

Charge Status

ON

steady

Charge in progress

ON

2 sec

OFF

.25 sec

ON

2 sec

(Repeats)

Approximately

80% charged

OFF

2 sec

ON

.25 sec

OFF

2 sec

(Repeats)

Charge completed

The battery pack may be kept on SKC-approved Li-Ion battery chargers for
an indefi nite time.

Cautions:

• Do not charge or operate pump from charger in hazardous

locations.

• Use only the charger and battery packs designed for the AirChek

XR5000 pump to ensure reliable performance. Failure to do so will

void any warranty.

• For safe operation in hazardous locations, ensure the pump label

contains the

logo and the battery pack label contains Cat.

No. P85004 or P85002. Use of any other battery pack or device to

power the pump voids the UL Listing for intrinsic safety.

• Tampering with the battery pack or using a repaired or rebuilt

battery pack voids any warranty and UL Listing for intrinsic safety.

• Do not open, disassemble, short circuit, crush, incinerate, or

expose the battery to fi re or high temperatures.

• Use only the SKC-approved charger for this pump. Use of an

unapproved charger may damage the battery and pump and voids

any warranty.

• Failure to follow warnings and cautions voids any warranty.

Setup