Maintenance, 17 resetting pump to manufacturer settings – SKC 210-5000 AirChek XR5000 User Manual
Page 20
17
Resetting Pump to Manufacturer Settings
If the pump does not operate as expected, perform the following procedure:
1. Remove the battery pack (see Removing Existing Battery
Pack).
2. On the pump keypad, press and hold and simul-
taneously while attaching the pump to the new battery
pack. The LCD should display the software version
number (525X).
Do not release hold on the two keys until the pump is fi rmly attached to
the battery pack.
3. Release
and .
4. Press
2 times. The LCD should read 0. If it does not, repeat Steps 1 through 4
until successful.
5. Install 2 screws and use a Phillips head screwdriver to tighten screws in an alter-
nating fashion.
Cautions:
• For safe operation in hazardous locations, ensure the pump label
contains the
logo and the battery pack label contains Cat.
No. P85004 or P85002. Use of any other battery pack or device to
power the pump voids the UL Listing for intrinsic safety.
• Do not charge or operate pump from charger in hazardous locations.
• Use only the charger and battery packs designed for the AirChek
XR5000 pump to ensure reliable performance. Failure to do so
voids any warranty.
• Use only SKC-approved parts to ensure reliable performance and
to maintain the UL Listing for intrinsic safety. Failure to do so voids
any warranty.
• Tampering with the battery pack or using a repaired or rebuilt
battery pack voids any warranty and UL Listing for intrinsic safety.
• Do not open, disassemble, short circuit, crush, incinerate, or
expose the battery to fi re or high temperatures.
• Use only the SKC-approved charger for this pump. Use of an unap-
proved charger may damage the battery and pump and voids any war-
ranty.
• Failure to follow warnings and cautions voids any warranty.
Maintenance
2