Note

Tip

A low battery fault may occur instead of a fl ow fault when there is a low battery
charge at the time of the fault, excessive back pressure, and/or when there is a
very short distance between the restriction and the pump inlet (e.g., fi nger fault
versus pinched tubing). The fl ow fault icon will not appear and auto-restart
will not be activated under these conditions.
A low battery fault icon (see
page 7) will appear instead and the pump will go to Sleep.

If pump goes to Sleep while in fl ow fault, the fl ow fault icon may remain on the
display when the pump is subsequently turned on. To remove the icon from the
display, place pump in Hold, press

 to enter user setup functions, and

press [

] when CLr appears.