Setup – SKC 210-5000 AirChek XR5000 User Manual
Page 13
10
During a run delay with continuous run:
a. The blue LED on top of the pump case will fl ash during the run delay even though
the pump is not running.
b. Run delay time remaining will count down in minutes and
display on the LCD.
c. Once the run delay time has elapsed, the pump will start
running. Accumulated run time will count up in minutes.
d. Run can be paused (Hold) by pressing []. The accumulated run time display
will freeze. Run can be resumed by pressing []. Accumulated run time dis-
play will resume.
e. Once the run is complete, stop the pump by pressing []. Accumulated run
time will display.
If pump goes to sleep following the continuous run and is awakened, the
pump will display accumulated run time.
Setting a Run Delay and Timed Run
With pump in a non-running state (no fl ashing blue LED):
1. Press to enter user setup functions.
2. Press until fl ashing Set Timed Run and min appear
on the display.
3. Press to increase or to decrease minutes (1 to
9999 minutes).
4. Press
to scroll to a fl ashing Set Run Delay and min.
5. Press
to increase or to decrease minutes (1 to
9999 minutes).
6. Press [] to accept run delay and timed run settings
and to exit user setup functions. The run delay setting
will appear on the LCD and the pump will be ready to start run delay.
7. Press [] to activate the pump.
During a run delay with timed run:
a. The blue LED on top of the pump case will fl ash dur-
ing the run delay even though the pump is not running.
b. Run delay time remaining will count down in minutes
and display on the LCD.
c. Once the run delay time has elapsed, the pump will
start running. “Timed Run Remaining” will display
and count down in minutes on the LCD.
d. Accumulated run time can be displayed by pressing
and holding .
e. Run can be paused (Hold) by pressing []. The time remaining and accumu-
lated run time displays will freeze. Run can be resumed by pressing []. Time
remaining and cumulative run time will resume
Setup
Note