CALIBRATION AND SAMPLING

Calibration

Calibrate pump fl ow rate with the sampling head loaded with a cartridge in
line. See pump and calibrator operating instructions.

Wear disposable, clean, lint-free nylon or powder-free surgical gloves to
handle the sorbent and cartridge.

Ensure pump has run for 5 minutes
before calibrating. Ensure rain
cover is removed from inlet and
that sampling head is completely
assembled with a fully loaded
cartridge (see Inserting a Cartridge
into the Sampling Head).

Thread calibration adapter into

sampling head inlet.

Set and calibrate pump fl ow rate to 10 L/min (see Leland

Legacy Quick Guide on page 9). Record the pre-sample fl ow

rate. See pump and calibrator operating instructions.

1

When calibration is completed,

disconnect calibrator and tubing

from calibration adapter. Remove

calibration adapter from sampling

head.

Use provided short length of calibration tubing to connect inlet of calibration

adapter to outlet of a calibrator to form a calibration train.

Unscrew quick-connect plug on side of case. Use tubing with quick-connect

fi tt ing to att ach case (pump) inlet to exhaust of sampling head.

Ensure O-ring is installed on the quick-connect fi tt ing before inserting it
into the inlet. Absence of the O-ring can aff ect measurements. See page 14
for Replacement Parts.

Inlet

Quick-

connect

plug

Inlet line

to pump

Quick-

connect

fi tting on

tubing

Inlet to

pump

Calibration adapter

Flowmeter

outlet

Sample

head

outlet