Leland legacy quick guide, Terms, Programming sequences – SKC DCS (Deployable Cartridge Sampler) System User Manual
Page 12
9
Leland Legacy Quick Guide
Terms »
Star button Ò
• Scrolls through run time data and Setup options
Up and down arrow buttons ST
• Toggle between display choices and increase or decrease sampling parameters in Setup
Button sequence
T
Ò= press buttons individually
[ST] = press simultaneously
ÒSTÒ
= security code, always press in sequence
Security code ÒSTÒ
• Prevents unauthorized changes to the pump’s sampling program
Programming Sequences »
• To activate pump (e.g., to change pump from Sleep to Hold):
Press any button.
• To change pump from Hold to Run or Run to Hold:
Press [ST].
• To reset accumulated data:
Press [ST], then ÒSTÒ. Press Ò until CLr displays then press [ST]; press Ò until End displays then
press [ST].
• To set pump fl ow rate:
Press [ST], then ÒSTÒ. Flow rate and SET fl ash. Press S or T to change fl ow rate. Press Ò until End
appears then press [ST] to save setting and place pump in Hold.
• To calibrate fl ow rate with standard calibrator:
Press [ST], then ÒSTÒ. Flow rate and SET fl ash. Press S or T to change fl ow rate. Press Ò once. ADJ
displays. Press S or T until desired fl ow rate is indicated on calibrator. When fi nished, press Ò until End displays
then press [ST] to save new setting and place pump in Hold. For CalChek Calibration, see operating instructions.
• To change temperature scale from F to C or C to F:
Press [ST], then ÒSTÒ. Press Ò until temperature displays. Press S or T to switch units; press Ò until
End displays then press [ST] to save new setting.
• To change atmospheric pressure scale (mm, mb, In):
Press [ST], then ÒSTÒ. Press Ò until pressure displays then press S or T to switch units; press Ò until
End displays then press [ST] to save new setting.
• To change time scale (12 Hr/24 Hr/Dela):
Press [ST], then ÒSTÒ. Press Ò until 12 Hr, 24 Hr, or Dela displays then press S or T to switch units;
press Ò until End displays then press [ST] to save new setting. To set delayed start (Dela), see operating
instructions.
• To change clock:
Press [ST], then ÒSTÒ. Press Ò until clock displays then press S or T to change fl ashing hour; press Ò to
move to minutes and S or T to change setting. Press Ò until End displays then press [ST] to save new setting.
• To change the sampling time function:
Press [ST], then ÒSTÒ. Press Ò until ST L/min displays then press S to change fl ashing digit; press Ò until
End displays then press [ST] to save new setting. To delete, follow above steps and press T until 0 appears.
Exit Setup.
Note: When in Setup, choosing Esc instead of End will exit Setup without saving new settings.