Gentec-EO M-LINK User Manual
Page 37

M-
LINK User’s Manual Revision 3.0
31
CUR
Query only. Returns the current measurement. The argument is 1 for Energy
Heads, Radiometric Heads, or Radiometer heads in Energy mode, and 0 for
all others.
Example
Send cur1CRLF
Reply DataCRLF
Data is in a hexadecimal data string. See the note below: Data Format used
by M-LINK for how to decode this value.
AVG
Sets or Queries the sampling size of the noise suppression.
Example
Send avg16CRLF sets sampling size to 16 pulses, M-LINK replies Ok.
avgCRLF queries the current average.
Example
avgCRLF
Reply 16CRLF
Valid averages are powers of 2 from 1 to 128. If a non valid argument is sent,
Err is returned and the value is left unchanged.
HVR
Query only. Returns the head version.
Example
Send hvrCRLF
Reply VersionCRLF
SSN
Query only. Returns the head serial number.
Example
Send ssnCRLF
Reply Serial NumberCRLF