Gentec-EO M-LINK User Manual
Page 36

M-
LINK User’s Manual Revision 3.0
30
POL
Sets the external trigger polarity for energy or radiometric measurement
Example
Send pol1CRLF Trigger polarity is rising edge.
Send pol0CRLF Trigger polarity is falling edge.
PRB
Query only. Returns the head connected flag. If no head is present the
command does not reply. This allows the control software to periodically poll
for a head connection.
Example
Send prbCRLF
Reply NCRLF
N = 1, head is connected. No reply means no head is connected.
AUT
Command only. Enables or Disables Auto Ranging.
Example
Send aut1CRLF Enable Auto Range. Replies Ok.
Send aut0CRLF Disables Auto Range, Replies Ok.
NAM
Query only. Returns the head name.
Example
Send nameCRLF
Reply Head NameCRLF
SSE
Command only. Enables or Disables Single Shot Energy for Thermopile
Heads.
Example
Send sse1CRLF Enable SSE Mode, M-LINK replies Ok.
Send sse0CRLF Disables SSE Mode, M-LINK replies Ok.
If a Thermopile head is not in use, M-LINK replies Thermopile Head Only.