Gentec-EO M-LINK User Manual
Page 34

M-
LINK User’s Manual Revision 3.0
28
STR
Starts or stops the data flow.
Example.
Send str1CRLF The M-LINK sends data to the com port. Data is sent at the
pulse rate for Energy, at the chop rate for Radiometers, and at 10Hz for
Power.
Send str0CRLF M-LINK stops sending data and replies Ok.
Data is the hexadecimal data string. See the note below for how to decode
this value.
MAX
Query only. Returns the maximum range index for the head in use.
Example
Send maxCRLF
Reply 27CRLF
MIN
Query only. Returns the minimum range index for the head in use.
Example
Send minCRLF
Reply 20CRLF