General Tools and Instruments WDCFM8912 User Manual

Page 12

background image

2. If the saved S value is different from your paper, press

M

HOLD

key

to move to the digit which you want to edit and then use

L

Mx/Mn

key to change the value.

3. Press

ENTER

key to save new input S value and then meter will

enter second value input automatically.
IMPORTANT NOTE: While setting the slope, only input 5 digits after
decimal point, so user need to round off the 6th digit.
User should also input two digits before decimal.
Example:
1. If the S value is 8.239678, user should input 08 23968
2. If the S value is 11.23968, user should input 11 23968
3. If the S value is 0.2396838, user should input 00 23968 (Fig. S)

(Fig. S)

INTERCEPT (VALUE Z)
1. While press

ENTER

key save the S setting, the meter will enter

intercept input automatically. The last digit will flash to indicate
the Z value is now ready for edit.

2. If the last saved Z value is different from your paper, press

M

HOLD

key to move to the digit which you want to edit and then use

L

Mx/Mn

key to change the value.

3. Press

ENTER

key to save new input Z value and then meter will be

back to normal RH value mode.

IMPORTANT NOTE: While setting Z value, only input 3 digits after
decimal point, so user need to round off the 4th digit.