Scale platform troubleshooting, Scale platform testing, Load cell testing – Fairbanks Aegis Drum Scales User Manual

Page 20: Zero, 25% of the load cell capacity

background image

Section 4: Service & Maintenance

04/10

20 51190 Rev. 2

4.2. SCALE PLATFORM TROUBLESHOOTING

Except for severe structural damages, most Platform Assembly problems can be
traced to the following causes.

Material under or around the Platform.

Broken Load Cell Feet.

Faulty Load Cells.

Incorrect, loose or damaged Load Cells.

4.2.1. Scale Platform Testing

1. Inspect the Interface Cable from the Platform to the Instrument for visible breaks

or cracks.

2.

ZERO

the Instrument Display.

3. Apply a test load of

25% of the Load Cell capacity

to one corner.

─ The Instrument should display a weight reading within 0.1% of the applied

weight, or One Instrument Division, whichever is greater.

4. Repeat Step 3 for all the corners, placing the same Test Load on each corner.

4.2.2. Load Cell Testing

When corners do not match the correct tolerances, unsolder each Load Cell Cable,
then test each Load Cell for the settings on the following chart.

TEST

READING

REMARKS

Green to Black (Input)

1106 Ohms (+5 / -2 Ohms)

Input Resistance

Red to White (Output)

1000 Ohms (+5 / -2 Ohms)

Output / Bridge Resistance

Yellow (Shield) to Load Cell Case

More than 1,000 megohms

Insulation Resistance

Input and Output Leads to Shield

Input and Output Leads to Case

If any Load Cell fails, it should be replaced according to Subsection 4.4.