To test, press any key other than the LOGO or ENTER keys; to resume normal operation

press the LOGO key.


4.

To perform a total memory reset press the ENTER key twice in succession.


AUTOMATIC DATA OUTPUT
(Function 7)

1.

Activate Special Functions.


2.

Press key 7 to activate the Data Output function.


3.

Press the up or down arrow as needed to cause the display to read the desired time

interval (002 to 999 seconds) between reports. Press ENTER. A report is a continuous

transmission, through the serial port, of all six data factors (measurements and

calculations).

4.

To stop data output, repeat steps 1 & 2 and set the time interval in step 3 to “000”

followed by the ENTER key.


To PRINT an instruction manual
(Key 8)

1.

Connect a serial printer to the instrument’s serial port. Turn on the printer and switch

it

“on-line”.


2.

Activate Special Functions, then press key 8.


3.

This abridged version of the instrument’s instruction manual will be printed.

-18-

To check the VERSION of your instrument’s software
(Key 9)

1.

Activate Special Functions.


2.

Press key number 9.


3.

The display will show the model number and the month on the first line, the hour and

the date on the second line, and the year on the third line. The model, month, day, and

year correspond to the date of software release.


4.

Press any key to resume normal operation.


ELECTRONIC ZEROING OF OXYGEN SECTION