Oxygen Monitor System

5 Failure and Error Codes

Teledyne Analytical Instruments

39

install a new sensor.

102

Bad zero and slope error
(both the zero and slope
are out-of-range)

See recommended actions
for errors 100 and 101
listed above.

203

Unstable span error (a
stable reading could not be
obtained during the span
cycle of a 1MAN
calibration)

Check the gas connections
for leaks and attempt to
perform another 1MAN
calibration. If this does not
correct the problem, install
a new sensor.

214

Unstable zero error (a
stable reading could not be
obtained during the zero
cycle of a 1MAN
calibration)

Check the gas connections
for leaks and attempt to
perform another 1MAN
calibration. If this does not
correct the problem, install
a new sensor.