3 Installation

Model 3220

Teledyne Analytical Instruments

20

3.6.2 Control Unit Jumper Settings

The Control Unit may be configured by changing the installation of

the jumpers on the PCB. See the following sections for complete
jumper installation information. The default jumper installation is
shown in the table below.

Table 3-1: Control Unit Jumper Configuration

Jumper Installation Function

JP1

Pins 1 & 2

Fail-safe Failure Alarm

JP2

Pins 1 & 2

Fail-safe Alarm 2

JP3

Pins 1 & 2

Fail-safe Alarm 1

JP4

Pins 1 & 2

Normally closed ext.
aud. Alarm

JP5

Pins 1 & 2

Fail-safe ext. aud.
Alarm

JP6

Pins 1 & 2

Ext. aud. contact
closures output

JP7

Pins 1 & 2

Ext. aud. contact
closures output

JP8 Thru JP16

OFF

RS-485 connections
not routed

JP17

ON

Internal aud. Alarm
enabled

JP18 & JP19

ON

Ground planes
connected


Three alarms (Alarm 1, Alarm 2, and Fail) may be configured for

either fail-safe (default) or non fail-safe operation according to the table
below:

Table 3-2: Fail-safe/Non Fail-safe Alarm Configuration

Alarm Jumper Failsafe

Non-failsafe

Alarm 1

JP3

Pins 1 & 2

Pins 2 & 3

Alarm 2

JP2

Pins 1 & 2

Pins 2 & 3

Failure

JP1

Pins 1 & 2

Pins 2 & 3