Model No. CI-6605A
Using the Temperature Sensor with a pH, Dissolved Oxygen, or Conductivity
3
Sulfuric Acid. For more information about chemical compatibility with
#304 stainless steel, see the Cole-Parmer web site
(www.coleparmer.com/techinfo).
Using the Temperature Sensor
with a pH, Dissolved Oxygen,
or Conductivity Sensor
The Stainless Steel Temperature Sensor is electrically
grounded. Use an optional Teflon sleeve to isolate the probe
of the Stainless Steel Temperature Sensor when it is used with
electrically susceptible sensors, such as the pH (CI-6507A)
and Conductivity Sensor (CI-6729)
Using the Temperature Probe
in Air-Tight Containers
The diameter of the sensor’s stainless steel probe is slightly
smaller than the 1/4-inch hole found in many rubber and cork
stoppers. If an airtight seal in a stopper is required, the diame-
ter of the stainless steel probe must be increased. A 6-inch
piece of plastic tubing is included with the Stainless Steel
Temperature Sensor for this purpose.
The tubing may be trimmed as required. A little bit of glycer-
ine may be used on the tubing to assist in the process of slip-
ping the tubing over the probe. For the best temperature
response, place the tubing and stopper as close to the probe
handle as is practical.
Figure 3: Probe with stopper and tubing
Specifications
Technical Support
For assistance with any PASCO product, contact PASCO at:
Limited Warranty
For a description of the product warranty, see the PASCO catalog.
Copyright
The PASCO scientific 012-08060C Stainless Steel Temperature Sensor
Instruction Sheet is copyrighted with all rights reserved. Permission is
granted to non-profit educational institutions for reproduction of any part of
this manual, providing the reproductions are used only in their laboratories
and classrooms, and are not sold for profit. Reproduction under any other
circumstances, without the written consent of PASCO scientific, is prohib-
ited.
Trademarks
PASCO, PASCO scientific, DataStudio, PASPORT, and ScienceWorkshop
are trademarks or registered trademarks of PASCO scientific, in the
United States and/or in other countries. All other brands, products, or ser-
vice names are or may be trademarks or service marks of, and are used to
identify, products or services of, their respective owners. For more infor-
mation visit www.pasco.com/legal.
Temperature range
-35ºC to +135ºC,
-31ºF to +275ºF
238 to 408 K
Accuracy
±0.5ºC or better
Resolution
0.05ºC
Probe composition
#304 stainless steel
Address: PASCO scientific
10101 Foothills Blvd.
Roseville, CA 95747-7100
Phone:
916-786-3800 (worldwide)
800-772-8700 (U.S)
Fax:
(916) 786-3292
Web:
www.pasco.com