S p e c i f i c a t i o n s

3

For step 5, put the probe and thermometer into hot water
(boiling hot, if available). When the thermometer
reaches its highest point, enter the thermometer reading
in the “Standard Value” text box. Click “Set Current
Value to Standard Value” and click “Next”.

In the final step, review your calibration and click “Fin-
ish”.

Using SPARKvue Software

After SPARKvue starts, the Sensor Parameter screen shows
the measurements that the sensor can make.

Touch “Temperature” and then touch “Show” to open a
page with a graph display.

In the graph page, touch the “SPARKvue Help” icon
(

) to open the SPARKvue User’s Guide page.

Touch “Setting up an experiment” and then touch “Cali-
brating sensors”.

Touch “Performing a calibration” and then touch “Per-
forming a 2-point calibration”.

Put the probe and the thermometer in ice water. Follow
the instructions in the SPARKvue Help page about “Cal-
ibration Point 1”. (Use the keypad window that opens to
enter the thermometer reading and then touch “OK”).

Put the probe and thermometer into hot water (boiling
hot, if available). Follow the instructions in the SPARK-
vue Help page about “Calibration Point 2”.

When finished, touch “OK” until you return to the graph
page.

Using Xplorer GLX Datalogger

Refer to the Xplorer GLX Datalogger User’s Guide for
calibration instructions.

Specifications

More Information

For the latest information about the Temperature Sensor,
visit the PASCO web site at www.pasco.com and enter
“PS-2125” in the Search window.

Technical Support

For assistance with any PASCO product, contact PASCO at:

Limited Warranty For a description of the product warranty, see the
PASCO catalog. Copyright The PASCO scientific Instruction Sheet
is copyrighted with all rights reserved. Permission is granted to
non-profit educational institutions for reproduction of any part of this
manual, providing the reproductions are used only in their laborato-
ries and classrooms, and are not sold for profit. Reproduction under
any other circumstances, without the written consent of PASCO sci-
entific, is prohibited. Trademarks PASCO, PASCO Capstone,
PASPORT, SPARK Science Learning System, SPARK SLS, and
SPARKvue are trademarks or registered trademarks of PASCO sci-
entific, in the United States and/or in other countries. For more infor-
mation visit www.pasco.com/legal.

Product End of Life Disposal Instructions:

This electronic product is subject to disposal and recycling
regulations that vary by country and region. It is your
responsibility to recycle your electronic equipment per your
local environmental laws and regulations to ensure that it
will be recycled in a manner that protects human health and
the environment. To find out where you can drop off your
waste equipment for recycling, please contact your local
waste recycle/disposal service, or the place where you pur-
chased the product.

The European Union WEEE (Waste Elec-
tronic and Electrical Equipment) symbol
(to the right) and on the product or its
packaging indicates that this product must
not be disposed of in a standard waste con-
tainer.

Item

Value

Range:

-35 to 135°C

Accuracy:

±0.5°C

Resolution:

0.01°C or better

Response TIme

Wait 15 seconds for stable
readings in liquids, and wait
30 to 60 seconds for stable
readings in gases.