beautypg.com

Specifications technical support, One-point and two-point calibrations, Over-sampling – PASCO PS-2181 Dual Pressure Sensor User Manual

Page 3

background image

®

D u a l P r e s s u r e S e n s o r

P S - 2 1 8 1

3

One-point and Two-point Calibrations

Though it is usually not necessary, a one-point or two-point cali-
bration can be performed on any of the measurements to make
them more accurate. To do so, you must have a very accurate,
independent means of measuring pressure such as a barometer.
For step-by-step instructions refer to the documentation for
DataStudio, the Xplorer, or the Xplorer GLX.

Over-sampling

The Dual Pressure Sensor uses dynamic variable over-sampling
to reduce noise, produce smoother data, and improve the mea-
surement resolution. This effect is especially noticeable when
very small pressure changes or differences are measured. The
degree of dynamic variable over-sampling that takes place within
the sensor depends on the sample rate. To maximize the
over-sampling, set the sample rate as low as possible for a given
application. Maximum over-sampling occurs at sample rates of
1 Hz or slower.

Specifications

Technical Support

For assistance with any PASCO product, contact PASCO at:

Limited Warranty

For a description of the product warranty, see the

PASCO catalog.
Copyright

The PASCO scientific 012-09969A

Dual Pressure Sensor

Instruction Sheet is copyrighted with all rights reserved. Permission is
granted to non-profit educational institutions for reproduction of any part of
this manual, providing the reproductions are used only in their laboratories
and classrooms, and are not sold for profit. Reproduction under any other
circumstances, without the written consent of PASCO scientific, is prohib-
ited.
Trademarks

PASCO, PASCO scientific, DataStudio, PASPORT, Xplorer,

and Xplorer GLX are trademarks or registered trademarks of PASCO sci-
entific, in the United States and/or in other countries. All other brands,
products, or service names are or may be trademarks or service marks of,
and are used to identify, products or services of, their respective owners.
For more information visit www.pasco.com/legal.

Select Dual Pressure Sensor

Select Calibrate all...

Select 1 Point (Adjust Offset Only)

Click Read From Sensor

Click OK

Range

Absolute pressure: 0 to 200 kPa
Differential pressure:

−100 kPa to +100 kPa

Resolution

0.01 kPa at 10 Hz

Repeatability

1 kPa

Units of Measure

kPa, N/m

2

, psi

Max. Sample Rate

1000 Hz

Included tubing

Material: polyurethane
Inside diameter: 3.2 mm (

inch)

Length: 2.4 m (8 feet)

Address: PASCO scientific

10101 Foothills Blvd.
Roseville, CA 95747-7100

Phone:

916-786-3800 (worldwide)
800-772-8700 (U.S.)

Fax:

(916) 786-7565

Web:

www.pasco.com

Email: