13 TROUBLE-SHOOTING

PROBLEM PROBABLE

CAUSE

REMEDIAL

ACTION

Nothing is displayed when
the ON/OFF key is
selected.

1. Batteries not in place.

2. Batteries (+ & - poles) not

in correct polarity.

3. Weak

batteries.

1. Insert

batteries.

2. Re-insert batteries in the

correct polarity.

3. Replace batteries or

attach AC adapter.

Unstable reading

1. Insufficient reference

electrolyte in electrode.

2. Broken

electrode.

3. External “noises” or

induction (e.g. electrical
‘noise’ caused by a
nearby motor).

4. Dirty

electrode.

1. Fill electrode with

reference electrolyte.

2. Replace

electrode.

3. Move or switch off

interfering device.

4. Clean the electrode.

Rejuvenate if necessary.
Recalibrate the meter
with the cleaned probe.

Slow responses.

1. Dirty/Oily electrode.

1. Clean electrode.

Rejuvenate if necessary.

Meter is not responding to
key press.

1. HOLD mode in operation.

2. Internal program error.

1. Cancel HOLD mode.

2. Reset all internal

programs by re-inserting
battery.